background image

 

 

 

©  PCE Instruments 

 

Operating instructions 

 When using the meter for the first time, the sensor  of the oxygen meter must be filled 

with electrolyte solution OXEL-03 and then calibrated. 

 

3.1 

Calibration 

Before the measurement, the PCE-DOM 10/20 must be calibrated in 

fresh air. 

First remove the 

grey protective cap from the sensor. Then switch on the test instrument using the on/off key. The 
display then shows a measured value and the current temperature:  

 

The upper, large display shows the current measured value. Wait approx. 3 minutes until the 
display has stabilised and the measured value no longer fluctuates. 
 

 

Содержание PCE-DOM 10

Страница 1: ...es Oxygen Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...ns 4 3 1 Calibration 4 3 2 Measurement of dissolved oxygen in liquids 5 3 3 Measurement of atmospheric oxygen 6 3 4 Temperature measurement 6 3 5 Freezing data in the display 6 3 6 Save measured data...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...DOM 20 4 m Temperature units C F Display LC display 29 x 28 mm Temperature compensation automatically Memory MIN MAX Automatic power off after about 15 minutes Operating conditions 0 50 C 80 RH Power...

Страница 5: ...Battery compartment 3 7 Protection cap 2 2 2 PCE DOM 20 3 1 Display 3 2 On Off key 3 3 HOLD key 3 4 REC key 3 5 Sensor with diaphragm 3 6 Battery compartment 3 7 Sensor connection 3 8 Sensor plug 3 9...

Страница 6: ...ore the measurement the PCE DOM 10 20 must be calibrated in fresh air First remove the grey protective cap from the sensor Then switch on the test instrument using the on off key The display then show...

Страница 7: ...r the calibration has been carried out as described in chapter 3 1 the oxygen meter can be used to measure dissolved oxygen in liquids Press the UNIT key for three seconds to change the unit from O2 t...

Страница 8: ...nsures that after activating this function the minimum and maximum measured values are saved in the display 3 6 1 Save maximum value Press and release the REC key Then the REC icon appears in the disp...

Страница 9: ...can no longer be calibrated or the reading does not appear stable on the display the following points should be considered 4 2 1 Testing the electrolyte Check the condition of the electrolyte in the...

Страница 10: ...filling the electrolyte Remove the black cap with the diaphragm from the sensor head Clean the sensor with tap water Fill new electrolyte fluid into the new cap with the diaphragm OXHD 04 Then screw t...

Страница 11: ...pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devic...

Страница 12: ...ce stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments c...

Отзывы: