Function setting 

6.1 

Freeze measured value 

 

Press the HOLD kry to freeze the current readings on the display. HOLD" appears in 
the display. 

 

Press again to return to measuring mode. 

6.2 

MIN / MAX 

 

Press  the  MAX  /  MIN  key  to  activate  the  max/min  memory.  The  display  shows  the 
current and the maximum measured value. 

 

Press  the  MAX  /  MIN  key  again.  The  display  shows  the  current  and  the  minimum 
measured value. 

 

Press  the  MAX  /  MIN  key  again.  The  display  shows  "Max  /  Min"  and  the  current 
measured  value  (only  available  for  the  measuring  functions  temperature  and 
frequency). The min and max values are saved in the background. 

 

Press and hold the MAX / MIN key to exit the function. 

6.3 

Hold peak value 

With the hold peak function, you can display the max. and min. value of the sine wave in the V 
AC measuring function. 

 

Press and hold the RANGE key to switch the function on or off. 

6.4 

Set measurement range 

You  can  influence  the  measurement  ranges  for  the  measuring  functions  voltage,  resistance, 
capacitance, frequency and µA. The default setting is automatic measurement range change. 

 

Press the RANGE key repeatedly until the appropriate measurement range is set. 

 

Press and hold the RANGE key to return to automatic range selection. 

6.5 

Display backlight 

Press and hold the HOLD key to turn the backlight on or off. 
(Depending  on  the  firmware  version,  the  display  backlight  switches  off  automatically  after  30 
seconds.) 

6.6 

Measuring spot illumination 

Press and hold the ZERO key to switch the measuring spot illumination on or off. 

6.7 

Bluetooth connection 

Press and hold the MODE key to switch the Bluetooth connection on or off. 
For use on the PC, download the PCE-DC 50 software from our download area: 

https://www.pce-instruments.com/english/download-win_4.htm

  

 
Use the dongle from the delivery scope as the Bluetooth interface. 
 
The "Meterbox Pro" app is available in app stores for connection to mobile devices. 
 

 

Содержание PCE-DC 50

Страница 1: ...DC Clamp Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...ty and diode test 8 5 7 Capacitance measurement 9 5 8 Frequency and duty cycle 9 5 9 A current measurement AC DC 9 5 10 Temperature measurement 9 6 Function setting 10 6 1 Freeze measured value 10 6 2...

Страница 3: ...l as diode tests if applicable may only be carried out in a voltage free state Before replacing the batteries or the fuses all test leads must be removed from the meter The case should only be opened...

Страница 4: ...t measurement AC DC test leads Measurement range Resolution Accuracy 0 00 500 00 A 0 01 A DC 1 0 6 digits AC 1 5 30 digits 0 0 5000 0 A 0 1 A DC 1 0 6 digits AC 1 5 30 digits Voltage measurement AC DC...

Страница 5: ...0 0 5000 0 nF 0 1 nF 3 5 10 digits 0 00 50 00 F 0 01 nF 3 5 10 digits 0 0 500 0 F 0 1 F 3 5 10 digits 0 000 5 000 mF 0 001 mF 2 0 10 digits Frequency measurement Measurement range Resolution Accuracy...

Страница 6: ...fast acting AC bandwidth AC A AC V 50 Hz 400 Hz AC measurement True RMS Operating conditions 5 C 40 C 41 104 F max 80 RH at 31 C Storage conditions 20 C 60 C 4 140 F max 80 RH Power supply 9 V block b...

Страница 7: ...er current clamp Indicator non contact voltage detection HOLD key display backlight ZERO key illumination MODE Bluetooth key Function rotary switch Display MAX MIN key Sockets for K type thermocouple...

Страница 8: ...rect current Duty cycle AC Alternating current F and C Temperature unit MAX Maximum value T1 T2 T1 T2 Temperature input 1 and 2 difference MIN Minimum value n m M k Measured value sign Battery low Con...

Страница 9: ...tip along the conductor to exclude false measurements due to twisted conductors If voltage is present the indicator for non contact voltage detection permanently glows in red 5 4 Current measurement A...

Страница 10: ...the cover of the input sockets upwards Set the rotary switch to position Press the MODE key repeatedly until the continuity test icon appears on the display Plug the test leads into the measuring inpu...

Страница 11: ...ottom 5 9 A current measurement AC DC Slide the cover of the input sockets upwards Set the rotary switch to position A Use the MODE key to select AC or DC Plug the test leads into the measuring input...

Страница 12: ...ch the function on or off 6 4 Set measurement range You can influence the measurement ranges for the measuring functions voltage resistance capacitance frequency and A The default setting is automatic...

Страница 13: ...ing Remove the test leads or temperature probes from the meter If you do not use the instrument for more than 60 days remove the batteries 7 1 Cleaning Wipe the meter regularly with a damp cloth 7 2 B...

Страница 14: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Страница 15: ...Unit 11 Southpoint Business Park Via Pesciatina 878 B Interno 6 Halkal Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul Unit...

Отзывы: