5.6. Display Setting
Display setting includes the function of turn-on and off for
“Color Offset”, “Tolerance Prompt”,
and “Screen Inversion”.
Click
“Color Offset” and turn on it, then it will display color offset prompt on the right of
△
L
、
△
a*
、
△
b* in the sample measurement interface.
Click
“Tolerance Prompt” and turn on it, then it will display “fail” or “pass” on the right of
△
E in
the sample measurement interface.
When
△
E is less than the value
of tolerance (setting in “Tolerance Setting”), it displays
“Pass”. Otherwise, it will display “Fail”.
“Color Offset” and “Tolerance Prompt:” as shown in Figure 41.
Figure 41
“Color Offset” and “Tolerance Prompt”
Click
“Screen Inversion” and turn on it, then all the display interfaces will be inversion
display.