PCE Instruments PCE-779N Скачать руководство пользователя страница 6

 

Manual 

 

 

6  LCD error messages 

Error message 

Reason 

 

“Hi” or “Low” is displayed when the temperature 
measured is outside the HAL and LAL limits. 

   

 

“Er 2” is displayed when the ambient temperature 
changes quickly. “Er 3” is indicated when the 
ambient temperature falls below 0 °C (32 °F) or 
e50 °C (122 °F). Allow at least 30 minutes 
for the thermometer to adapt to the room 
temperature. 

 

“Er 5” … “Er 9” means that you must reset the 
thermometer. To do so, turn the device off, remove 
the battery and wait for at least one minute before 
you reinsert the battery and turn the device back 
on. If the error remains, please contact PCE 
Instruments. 

 

 

 

“Hi” or “Lo” is indicated when the temperature 
measured is outside the measurement range. 

 

7  Storage and cleaning 

The device should be stored at room temperature. The sensor lens is the most sensitive part of the 
thermometer. Therefore, it should always be kept clean. Please be careful when cleaning it. Use a soft 
cloth or cotton swab with water or medical alcohol. Do not submerge any part of the instrument. 

8  Batteries 

The battery level indication can be as follows: 
 

 

“Battery OK”: measurements are 
possible. 

 

“Battery low”: measurements are 
still possible, but battery needs 
to be replaced. 

 

“Battery flat”: measurements are 
no longer possible. 

 

 

When the “Battery low” icon appears, the battery should immediately be replaced by a new 1.5 V 

AAA battery. To avoid malfunction, please switch off the device before replacing the battery. 
 

 Dispose of used batteries immediately and keep them away from children. 

 
For the disposal of batteries, the 2006/66/EC

 

directive of the European Parliament applies. Due to the 

contained pollutants, batteries must not be disposed of as household waste. They must be given to 
collection points designed for that purpose. 
 
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or 
give them to a recycling company which disposes of the devices in line with law. 

Содержание PCE-779N

Страница 1: ...Jupiter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Bus...

Страница 2: ...fety notes 3 3 Specifications 4 4 System description 4 5 Instructions 5 5 1 How to make a measurement 5 5 2 Functions 5 5 3 Settings 5 6 LCD error messages 6 7 Storage and cleaning 6 8 Batteries 6 9 C...

Страница 3: ...when the device is in use Be very careful when using the laser Do not point it towards people s or animals eyes The device may only be used in the approved temperature range The measurement range onl...

Страница 4: ...Note under an electromagnetic field strength of 3V m from 200 to 600 MHz the maximum error is 10 C 18 F Emissivity 0 95 default adjustable 0 1 to 1 in steps of 0 01 Response time T90 1 second Distance...

Страница 5: ...en the reading exceeds the high alarm HAL or falls below the low alarm LAL the high or low icon will flash and you will hear a beep sound Use the thermocouple socket 5 to connect the probe Target the...

Страница 6: ...ays be kept clean Please be careful when cleaning it Use a soft cloth or cotton swab with water or medical alcohol Do not submerge any part of the instrument 8 Batteries The battery level indication c...

Страница 7: ...contact PCE Instruments 9 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone 02380 987 035 9 2 P...

Отзывы: