When the battery voltage is under proper operation requirement, the            symbol will show on the 
LCD and the battery need to be replaced with new one. 

4.12 Digital Output: 

The Digital Output is a 9600bps N 81 serial interface. 

 
 

 

 

VI. Calibration Procedure 

 

Humidity Calibration: 

 

1. Turn the unit off. Press and hold 

MAX/MIN

HOLD

 and 

REC

 keys. While these three keys are 

pressed, turn the power on. Release these keys and all the segments on the LCD will blink. 

2. After step one, within 3 seconds, the user have to press 

°C/°F

 and 

TIME

 key at the same time 

to enter the calibration mode or the tester will go back to normal operation mode. 

3. When the tester get into calibration mode, the humidity reading will blink and “CAL1” will appear 

on the second display.   

4. Insert the humidity probe into the standard humidity cavity of 32.8%RH@25°C. Wait until the 

system to stabilize for 20 minutes then press 

MAX/MIN

 button to create the calibration data. If 

the unit recognizes the value is out of tolerance, the unit will sound 2 beeps and still remain at 
"CAL1”mode. If the tester recognizes the value is within tolerance, new calibration data is 
created and the tester will go into “CAL2” mode, which will be indicated by “CAL2” at 2

nd

 display. 

5. Insert the humidity probe into the standard humidity cavity of 75.3%RH@25°C. Wait until the 

system to stabilize for 20 minutes then press 

MAX/MIN

 button to create the calibration data. If 

the unit recognizes the value is out of tolerance, the unit will sound 2 beeps and still remain at 
"CAL2”mode. If the tester recognizes the value is within tolerance, new calibration data will be 
written into the memory and the calibration is done. 

 

Remark: 

1. When the user perform the humidity calibration, the environment should be kept at the stable 

condition (i.e. Constant temperature and constant humidity in the lab.) to in increase the 
accuracy. 

2. After the probe insert into the standard humidity cavity, the operator should wait at least 20 

minutes to let the condition in the cavity to stabilize. 

3. During the calibration, if the user press 

POWER

 button at any time, the tester will go back to 

normal operation mode and no calibration data will be changed. 

4. During the calibration mode, the user can restore the factory default value by press the 

HOLD

 

and 

REC

 buttons at the same time. 

5. Because it takes some time to stabilize the system, we recommend the operator first insert the 

probe into the 32.8%RH@25°C standard cavity and wait for at least 20 minutes, then power the 
unit on and start the calibration process. 

6. During the calibration, all the displayed reading is calibrated with the old calibration data and 

the auto power function is disabled. Until the calibration process is done, the tester will enable 
auto power function again. 

7. During the calibration, the temperature is fixed at °F scale and it is not selectable. 

 

Temperature Calibration: 

 

1. Turn the unit off. Press and hold 

MAX/MIN

HOLD

 and 

REC

 keys. While these three keys are 

pressed, turn the power on. Release these keys and all the segments on the LCD will blink. 

2. After step one, within 3 seconds, the user have to press 

°C/°F

 and 

HOLD

 key at the same time 

to enter the calibration mode or the tester will go back to normal operation mode. 

3. When the tester get into calibration mode, the temperature reading will blink and “CAL1” will 

Содержание PCE-313

Страница 1: ...9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF...

Страница 2: ...nd Temperature Measurement 3 4 3 Selecting the Temperature Scale 3 4 4 Data Hold Operation 3 4 5 DataLogger 3 4 6 Clock Setup 3 4 7 Recording Interval Setup 4 4 8 Time Operation 4 4 9 MAX MIN Operatio...

Страница 3: ...tions manual contains general product information and specifications Its internal memory can keep up to 16300 readings note1 It uses a RS232 interface to perform bi directional communication with PC I...

Страница 4: ...power off is enabled This indicates that the display data is being held m d month and day h m hour and minute m s minute and second Y year The Battery is not sufficient for proper operation REC This...

Страница 5: ...off last time 4 4 Data Hold Operation The user may hold the present reading and keep it on the display by pressing the HOLD button When the held data is no longer needed one may release the data hold...

Страница 6: ...pt in the memory simultaneously and updated with every new data point When the MAX symbol is display the Maximum is shown on the display Press MAX MIN again then the MIN symbol is on the display and t...

Страница 7: ...ue is within tolerance new calibration data will be written into the memory and the calibration is done Remark 1 When the user perform the humidity calibration the environment should be kept at the st...

Страница 8: ...utes to stabilize the system 2 During the calibration the user can leave the process by pressing the POWER button at any time and the calibration data will be kept unchanged 3 During the calibration m...

Страница 9: ...fore installing TestLink 2 Insert the TestLink CD ROM into your CD drive The TestLink installer should start automatically If it does not you can start it by running SETUP EXE from the root drive of t...

Страница 10: ...nd how many bytes have been received When data is loaded successfully there will be three new window show up see 5 2 Data Sets Window Display how many data sets were loaded and the detail information...

Отзывы: