2c. Enter a value less than 1
In the mV/V functions, the standard way to enter a value less than 1 is to press the leading 0
before pressing the decimal point. Though the decimal point can be entered first, the decimal
point will not be shown in the LCD.
3. Hz, Frequency Output
1. Turn the power on, then plug the test leads into the output connectors of calibrator
accordingly (Black to black, red to red). Attach alligator clip if necessary.
2. Move the sliding switch to Hz position
3. Press the keypad (excluding the decimal point) to enter the value of Hz directly.
4. Using the test leads or alligator clips, touch or clip on the terminals to be calibrated.
5.Because not all the frequencies are available between 126 to 62500Hz, calibrator will
automatically adjust and display the users input value to a frequency, which is available and
always larger or equal to users input.
Note: The resolution of Hz function is 1 Hz. For the range 1-125Hz, all the frequencies in
between are available. But for the range 126 - 62500Hz, not all the frequencies are available
(total 604 frequencies available). Please refer to section III Electrical Specifications for available
frequencies.