12
2. mV, V Output
2a. General Operation 0 - 100mV
1. Turn the power on, and wait until the STBY symbol disappears (about 2
min.).
2. Plug the test leads into the output connectors of calibrator accordingly
(Black to black, red to red). Attach alligator clip if necessary.
3. Move the sliding switch to mV, V position
4. Press the keypad (including the decimal point) to enter the value of mA
directly.
5. Using the test leads or alligator clips, touch or clip on the terminals to be calibrated.
Note: Always wait until STBY (standby) symbol in the LCD to disappear before
any operation.
Note: Maximum 5 digits can be entered. If users enter less than 5 digits (1 to 4 digits), users
must press ENTER button to indicate the end of entry. If users enter 5 or more digits,
calibrator will automatically end the entry and output specified value current.