4
Setup Menu
PCE-PH 28L pocket pH meter contains an integrated setup menu that is used to customize the displayed option to meet measurement
requirements.
MENU
DESCRIPTION
OPTIONS
DEFAULT
Set the pH buffer group for calibration and auto-
recognition.
USA (pH4.01/7.00/10.01)
USA
NIST (pH4.01/6.86/9.18)
Set the number of calibration points.
1 point
2 points
2 points
3 points
Set the default temperature unit.
Degrees Celsius
°C
Degrees Fahrenheit
Calibrate the temperature.
---
---
When the option is enabled, the meter will
automatically sense a stable reading and lock the
measurements.
Enable
Disable
Disable
When the option is enabled, the meter will
automatically switch off if no key is pressed within 8
minutes.
Enable
Disable
Disable
Reset the meter to factory default settings.
Enable
Disable
Disable
Setting the default option
1.
Press and hold the
Cal
key for 3 seconds to enter the setup menu, the meter shows a menu item and default option.
2.
If necessary, press the
Cal
key again to select the desired option.
3.
Press the
Enter
key to confirm, the meter moves to the next menu item.
4.
Repeat the steps above until the meter returns to the measurement mode. Setting is completed.
•
During the setting process, press the
Meas
key, the meter will exit the setup menu and return to the measurement mode.
•
If you do not want to calibrate the temperature, press the
Enter
key to skip the °C/CAL or °F/CAL option.
•
The Reset function will restore the meter back to factory default settings. If enabled, all of the calibration data and selected parameters
will be lost or reset, the meter must be recalibrated.
Option