10
Troubleshooting
LCD DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode dried out
Soak the electrode in 3M KCL solution at least 20 minutes.
Measured value is out of range
Check the electrode whether clogged, dirty or broken.
Incorrect pH buffer solutions
Using the fresh pH buffer solutions for calibration.
Electrode is broken
Replace the electrode.
pH Electrode Selection Guide
PCE-PH 28L pocket pH meter installed with a long-reach pH electrode that is used to measure the pH of the small volume samples. If this
electrode can not meet your measurement requirements, please refer to the table below to select an applicable sensor.
ELECTRODE
ORDER CODE
DESCRIPTION
General purpose pH electrode, suitable for measuring the non-
viscous, non-corrosive liquids.
Flat surface pH electrode, suitable for measuring the semisolid
substances, e.g., cheese, meats, paper, etc.
Specifications
pH
Model
PHscan30L
Range
-1.00~15.00pH
Accuracy
±0.01pH
Resolution
0.01pH
Calibration Points
1 to 3 points
pH Buffer Options
USA (pH4.01/7.00/10.01) or NIST (pH4.01/6.86/9.18)
Automatic Buffer Recognition
Yes
Temperature
Range
0~60°C, 32~140°F
Accuracy
±1°C
Resolution
0.1°C
Calibration Points
1 point
General
Temperature Compensation
0~60°C, 32~140°F, Automatic
Sensor Type
E-PHSCAN-LT-10K
Operating Temperature
0~60°C, 32~140°F
Relative Humidity
< 80%
Power Requirements
2 × 1.5V "AAA" batteries
Dimensions
240 (L) × 40 (Dia.) mm
Weight
100g