6. Operation 

Please make sure that all components listed below are included in your 
package. 

Qty  Description 

 

 

PCE-LDC 8 leak detector   

Sensor unit 

 

noise isolated 
headphones 

 

focus tube incl. focus tip   

Cable from instrument to 
sound probe 

 

Battery charger 

 

Transport case 

 

user manual 

 

Remark 

 

The sensor unit can be unplugged from the main instrument by 
pulling the unit out of the holder. A separate coiled extension cable 
is used to connect the sensor to the main unit. 

6.1 Operating Principle 

Air leaks produce wide-band ultrasounds in the range of 20... 80 kHz. 
The higher the frequency the more energy it contains. But higher 
frequencies cannot be transported in air that far. That is why the leak 
detector operates at a centre frequency of 40 kHz which compromises 
an optimum between energy and distance. Frequencies below and 
above are cut in order to minimize the noise level. 

 

Содержание PCE-LDC 8

Страница 1: ...English User Manual PCE LDC 8 Leak detector...

Страница 2: ...Should the device be tampered with in any manner other than a procedure which is described and specified in the manual the warranty is cancelled and the manufacturer is exempt from liability The devic...

Страница 3: ...3 Features 6 4 Technical Data 7 4 1 General 7 4 2 Electrical Data 7 4 3 Performance chart 7 5 Dimensional drawing 8 6 Operating 9 6 1 Operating Principle 9 6 2 Operating Procedure 10 6 3 Electrical c...

Страница 4: ...g this manual or the product please contact the manufacturer WARNING Compressed air Any contact with quickly escaping air or bursting parts of the compressed air system can lead to serious injuries or...

Страница 5: ...gulations before during operation Remarks It is not allowed to disassemble the product ATTENTION Measurement values can be affected by malfunction The product must be used properly and frequently main...

Страница 6: ...k through small openings The integrated laser pointer helps to spot the leak from distance The PCE LDC 8 leak detector is not developed to be used in explosive areas For the use in explosive areas ple...

Страница 7: ...on About 4 hours with Laser pointer on Charging temperature 10 C 45 C Charging time Around 1 5 hours Material of the detector PC ABS Dimensions See dimensional drawing on the next page Display 3 colou...

Страница 8: ...5 Dimensional drawing...

Страница 9: ...main instrument by pulling the unit out of the holder A separate coiled extension cable is used to connect the sensor to the main unit 6 1 Operating Principle Air leaks produce wide band ultrasounds i...

Страница 10: ...red User can change the sensitivity with the wheel 3 To activate the laser please press the button which is shown on the left side 4 Point with the laser at an assumed leak The display will show the...

Страница 11: ...eries in the EU the 2006 66 EC directive of the European Parliament applies Due to the contained pollutants batteries must not be disposed of as household waste They must be given to collection points...