c) To exit the memory record function, just press the
" REC button " for at least 2 seconds. The display will
revert to the current reading.
4-6 Zero adjustment procedure
Due to drift of environment temperature value, battery
power change or, meter used for a long time or other
reasons. The display value may exist not zero value
( few digits ) in case of no signal into the " Vibration
Sensor ". General speaking those not zero value will not
effect the measurement typically. However if intend to
make the precision measurement, the following zero
adjustment procedures should be executed as :
1)Press the " Function buttion " ( 3-5, Fig. 1 )
to the " Acceleration " position.
2)No signal into the vibration sensor.
3)Press the " Zero button " ( 3-3, Fig. 1 ) continuously
at least 2 second, the display will return to
zero value with default.
4-7 Data Logger
The data logger function can save 1000-point
data for the vibration function.
The data logger procedures are as following :
a)Press the " REC Button " ( 3-4, Fig. 1 ) once to
start the Data Record function and there will be a
" REC. " symbol on the display.
15