70
Example: set the sound level meter mode.
02 01 43
4D 45 4D 31
03 37 0D 0A
Return: ACK.
02 01 06 03 06 0D 0A
MEM?: Query Measurement Mode Setting
Instruction
Parameters
Explanation
MEM
Query parameter: ?
ASCII
M
E
M
?
Hex
4DH
45H
4DH
3FH
Byte
1
1
1
1
Return
Return the measurement mode
Example: query the measurement mode.
02 01 43
4D 45 4D 3F
03 39 0D 0A
Returns: the current measurement mode is level meter mode.
02 01 41
31
03 70 0D 0A
CALp1: Set Calibration Level and Calibrate by Measurement
☆
Note:
When CAL instruction is correctly received by the sound level meter, two ACK will be
returned at the beginning and the end of the calibration (several seconds will be spent by the
calibration). In the calibration history, ending with symbol
M
indicate the record was calibrate
by the method of by Measurement.
Instruction
Parameters
Explanation
CAL
p1: Calibration level;
Range: 0~199.9;
Default: 93.8
ASCII
C
A
L
93.8
Hex
43H
41H
4CH
39H
,
33H
,
2EH
,
38H
Byte
1
1
1
1~5
Return
ACK / NAK
Example 1: set the calibration level as 94dB and calibrate by measurement.
02 01 43
43 41 4C 39 34
03 00 0D 0A
Return: ACK.
02 01 06 03 06 0D 0A
Return again after calibration finished: ACK
02 01 06 03 06 0D 0A
Example 2: set the calibration level as 113.8dB and calibrate by measurement.
PCE-428 / 430 / 432