50
meter will power on when proper power supply available. It’s suitable for integrate into
other system, especially in those cases where power failure, sound level meter can
power on automatically from power shutdown.
Boot & Auto Meas.:
Need to change hardware mode switch to
Boot
. After select this mode, sound level
meter will not only power on when proper power supply available, but also start
measurement. When sound level meter was integrated into other system, it will power on
and start measurement automatically from power failure.
Hardware Mode Switch:
The hardware mode switch located in the battery compartment. It’s easily to be found
after remove the batteries. Please select the switch to Boot or Normal by nib or tweezers.
☆
Note:
Static electricity sensitive area. Eliminate static electricity before
operation.
4.4.10 USB Mode
Menu of
USB Mode
can set the working mode when connect
sound level meter to computer by USB cable.
Always Ask
,
USB Disk Mode
and
Modem Mode
can be select.
Always Ask:
It always ask which mode should apply when connect to computer by USB. Please make
choose in time, otherwise computer could not recognize the sound level meter due to
PCE-428 / 430 / 432