S-TYPE LOAD CELL OPERATION MANUAL 

 

4

 

6.1.2 Hysteresis 

 

The  difference  between  the  ascending  and  descending 
measured readings at 40% of full scale is used to calculate the 
hysteresis value. 

6.1.3 Best Fit Output 

 

The  best  fit  calibration  second-order  equation  has  been 
calculated  from  the  calibration  data  by  the  method  of  least 
squares.  Deviation  between  measured  output  and  best-fit 
output  is  calculated  and  displayed  in  the  column  next  to  the 
best-fit  output  for  each  measurement  increment.  The 
deviations  (%  Full  Scale)  of  measured  outputs  from  the 
calculated best fit are tabulated for each measured reading. 

6.1.4 Strain Gage Measurements 

 
 
 
 
 
 
 
 
 
 

6.1.5 Shunt Calibration Standard Resistor 

 

All  S-type  load  cell  calibrations  use  a  60K  Ohm  (0.1%) 
precision  resistor  shunt  calibration  value  that  is  supplied  into 
the calibration report. 

6.1.6 Static Error Band (SEB) 

 

The  static  error  band  (SEB)  is  determined  by  the  maximum 
deviations of  the ascending and descending calibration points 
from  the  best  fit  straight  line  through  zero  output.  The  SEB 
includes the effects of nonlinearity, hysteresis, and non-return 
to minimum load. 
 

7.0 SHUNT CALIBRATION DESCRIPTION 

 
Shunt  calibration  is  used  to  simulate  a  known  tension  or 
compression  load  on  a  load  cell.  The  calibration  certificate 
will indicate which leg of the bridge to apply the shunt resistor 
to for both tension and compression load simulation. Typically 
tension is simulated by inserting the shunt resistor between the 
+P and +S connector leads. Compression loading is simulated 
by  inserting  the  shunt  resistor  between  the  +S  and  –P 
connector leads. 
 

7.1 Resistor Value 

S-type  load  cells  have  a  nominal  2.0  mV/V  full  scale  output. 
For  a  350  ohm  strain  gage  bridge  the  precision  shunt  resistor 
is, 60,000 ohms 

±

 0.1%, simulates an output of approximately 

73% of the  full scale output  for the load cell. The calibration 
values for each bridge are found on the calibration certificates 
supplied with each load cell. 
 

7.2 Shunt Calibration Process 

To perform the shunt calibration, use the following procedure: 
 

1.

 

Stabilize  all  forces  on  the  load  cell.    If  possible, 

remove all loads. 

  
2.

 

Allow  the  powered  sensor  and  signal  conditioner  to 

warm up for a minimum of 30 minutes  

 
3.

 

Set the load indicator display to read exactly 00.000. 

 
4.

 

Connect  the  shunt  resistor  to  the  terminals  specified 

in  the  calibration  certificate,  and  adjust  the  span  or 
gain until the display reads the force value stated on 
the certificate. 

 
5.

 

Repeat  steps  1-3  to  verify  that  a  valid  calibration 

setting has been obtained. 

 
6.

 

If  possible,  apply  a  known  load  to  the  measurement 

system to further verify that the calibration has been 
accurately set up. 

7.3 Estimating Shunt Resistor for a Given Load 

The  following  formula  can  be  used  to  estimate  the 
approximate  value  of  shunt  resistor  required  to  simulate  a 
mechanical load. 
 
 

R

cal

 = (25

0

 * R

b

) / (Output

FS

 * L

cal

)   

 
Where: 

R

cal

 = Shunt Resistor (K ohms) 

 

R

b

 = Bridge Resistance (ohms) 

 

Output

FS

 = Full Scale output of the load cell (mV/V) 

 

L

cal

 = Load to be simulated, % of Load Cell Capacity 

 
 

8.0 MAINTENANCE 

 
Routine  maintenance  of  the  S-type  load  cell  should  include 
cleaning  the  electrical  connectors,  housings,  and  mounting 
surfaces  with  solutions  and  techniques  that  will  not  harm  the 
physical  material  of  construction.    Make  sure  liquids  are  not 
allowed  to  migrate  into  devices  that  are  not  hermetically 
sealed.  Such devices should only be wiped with a damp cloth, 
and  never  be  submerged  or  have  liquids  poured  on  them.  
Never use a pressure washer on the load cells. 

Bridge Resistance: 

350 Ohm Nominal 

Excitation: 

+P(A) to –P(D) Ohms 

Signals: 

+S(B) to –S(C) Ohms 

Leakage to Ground: 

> 5k GOhm 

Bridge Unbalance: 

±

1.0% Full Scale 

Output: 

2.0 mV/V Nominal 

Maximum Voltage: 

20 VDC 

 

Table 5

 

– Strain Gage Measurements 

Содержание M1630-04C

Страница 1: ...stallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbl...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...erations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 3 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Stand...

Страница 7: ...STALLATION Loads applied to S type load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing...

Страница 8: ...or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded I...

Страница 9: ...ision shunt resistor is 60 000 ohms 0 1 simulates an output of approximately 73 of the full scale output for the load cell The calibration values for each bridge are found on the calibration certifica...

Страница 10: ...aboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point descending Additional data points are avail...

Страница 11: ...5 in 51 mm x 64 mm x 16 mm 7 Weight 0 23 lb 104 g Housing Material Aluminum Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 002 in 0 05 mm Electrical Connector 10...

Страница 12: ...MODEL M1630 SERIES S CELL JAM 6 28 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 46140 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduc...

Отзывы: