Model M1630-04C

STRAIN GAGE LOAD CELL

Installation and Operating Manual

For assistance with the operation of this product,contact: 

PCB Load & Torque, Inc.

Toll-free: 866-684-7107

24-hour SensorLine™: 716-684-0001

Fax: 248-888-8266

E-mail: [email protected] 

Web: www.pcbLoadTorque.com

Содержание M1630-04C

Страница 1: ...stallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbl...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...erations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 3 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Stand...

Страница 7: ...STALLATION Loads applied to S type load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing...

Страница 8: ...or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded I...

Страница 9: ...ision shunt resistor is 60 000 ohms 0 1 simulates an output of approximately 73 of the full scale output for the load cell The calibration values for each bridge are found on the calibration certifica...

Страница 10: ...aboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point descending Additional data points are avail...

Страница 11: ...5 in 51 mm x 64 mm x 16 mm 7 Weight 0 23 lb 104 g Housing Material Aluminum Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 002 in 0 05 mm Electrical Connector 10...

Страница 12: ...MODEL M1630 SERIES S CELL JAM 6 28 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 46140 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduc...