2.2  MEASUREMENT SYSTEM CHECKOUT AND/OR GAIN ADJUST 

 
The Model 401B04 is inserted in place of the ICP

®

 sensor in your measurement system.  Connect the 

input of the 401B04 to a signal generator and the output to the ICP

®

 sensor signal conditioner.  Also 

connect any additional amplifiers or recording devices that will be used in the test, as shown in the figure 
above.  After switching on all equipment and allowing sufficient warm-up time, apply a known signal 
through the system with the signal generator and adjust gains, ranges, zeroes, etc., as needed.   
 

2.3  CALIBRATION OF LONG LINES 

 
Model 401B04 can be used, in conjunction with an oscillator or step function generator, to simulate an 
ICP® sensor driving a long input cable between the sensor and the ICP sensor signal conditioner.  Refer 
to PCB “General Guide to ICP® Instrumentation” for more details on the procedure.   
 
 

3.  FREQUENCY RESPONSE CONSIDERATIONS 
 

When used with a signal generator for system checkouts or testing of long lines, low frequency response of 
the 401B04 is DC. 
 
High frequency response of the Model 401B04, when driving short cables with small (~1V) signals, is 1 
megahertz.  Refer to PCB “General Guide to ICP® Instrumentation.” if long cables are to be driven.  Also, 
consult the factory if less than 2 mA constant current will be used to power the 401B04. 

 
 
4.  MEASUREMENT AND REPAIR 
 

The miniature size precludes most maintenance.  If used in a very humid or dirty environment, signal 
degradation, usually in the form of reduced output or loss of low frequency response, may occur.  If the 
connectors become dirty, clean with Freon TF ®, Genesolv, or other suitable solvent that leaves no residue. 
 
 
 
Manual Number: 40858 
Manual Revision: A 
ECN Number:  38393 

Содержание RH401B04

Страница 1: ...lator Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mai...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...er to the acceptance test procedure for your particular PCB signal conditioner dual mode charge amplifier or digital peak meter model Acceptance test procedures are available for most PCB electronic p...

Страница 7: ...the ICP sensor signal conditioner Refer to PCB General Guide to ICP Instrumentation for more details on the procedure 3 FREQUENCY RESPONSE CONSIDERATIONS When used with a signal generator for system c...

Страница 8: ...Jack Size Length x Width x Height 2 2 in x 1 1 x 0 9 55 9 mm x 27 9 x 22 9 Weight 2 50 oz 70 9 gm All specifications are at room temperature unless otherwise specified In the interest of constant pro...

Страница 9: ......

Отзывы: