 FORCE SENSOR OPERATION MANUAL 

 

 10 

10.0   TROUBLESHOOTING   

 

When a PCB signal conditioner with any of the following 

indicators are used, turn the power on and observe the voltmeter 

(or LED’s) on the front panel.   

 

 

 

NORMAL OPERATION 

 

 

Output voltage moves from YELLOW to GREEN slowly until 

charging is complete.  AC coupled signal conditioners require 

sufficient time to charge the internal coupling capacitor.  Allow 

signal conditioner to charge for five (5)  discharge time 

constants for stable operation.  In most cases, this is just a few 

seconds. 

 

Note: Most PCB force sensors have an output bias of 8-14 

VDC.  Refer to the specification sheet in this manual for the 

bias range of the model you are using.  If you are using a low 

bias sensor, the indicator will be at the bottom end of the green 

portion of the dial indicator, and may even be in the red portion. 

This is the expected range and indicates proper operation. 

 

 

11.0   MAINTENANCE 

 

The sensor connector must be kept clean, especially if it is 

operating in a dusty and/or wet environment.  Because the force 

sensor is of welded construction, it should be returned to the 

factory for servicing in the event of serious malfunction. 

 

Observe the following precautions in using the sensor: 

 

A.

 

Do not exceed the maximum  load levels for the force 

sensor (see specification sheet).  

 

B.

 

Do not subject the sensor to temperatures exceeding that 

of the specification, normally 250°F (121°C). 

 

C.

 

Do not apply voltage to the sensor without current-

limiting diodes or other current protection. 

 

D.

 

Do not apply more than 20 mA of current to the force 

sensor. 

 

E.

 

When mounting the force sensor, observe installation 

procedures detailed in Section 3.0 and as outlined on the 

specific sensor Installation Drawing to avoid over-

torquing when mounting. 

 

F.

 

Do not apply more than 30 volts to the sensor. 

 

G.

 

Avoid metal-to-metal impacts

 during applications as this 

application produces high-frequency energy and ringing 

within the sensor which could damage the internal 

crystal(s) or ICP amplifier.  Electrical low-pass filtering 

or a mechanical damping material can help reduce such 

effects. 

 

H.

 

Do not spin the sensor onto the cable

.  This may fatigue 

the cable center pin, causing cable damage.  Always 

insert the cable pin into the sensor and tighten the 

knurled cable nut to the sensor.  

 

 

ICP

 is a registered trademark of PCB Piezotronics 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MANUAL NUMBER:  18218 

MANUAL REVISION

 J

 

ECN NUMBER: 

53133

 

Содержание M221B02

Страница 1: ...comp 100 lb tension 50 mV lb Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...tallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description o...

Страница 7: ...ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Inst...

Страница 8: ...simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the spe...

Страница 9: ...eturning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for com...

Страница 10: ...t facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surface...

Страница 11: ...he left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total for...

Страница 12: ...itor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accom...

Страница 13: ...hundreds of feet meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent...

Страница 14: ...ant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents...

Страница 15: ...urned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specificat...

Страница 16: ...2 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side Mounting Thread 1 4 28 Female 1 4 28 Female All specifications are at room temperature unless otherwise specified In the inte...

Страница 17: ......

Отзывы: