 FORCE SENSOR OPERATION MANUAL 

 

 8 

5.0   OPERATION 

 

APPLICATION OF A FORCE

 

For best results, the applied force should be distributed evenly 

over the contact surface of the sensor.  Care should be taken to 

limit the bending moment induced into the sensor  by edge 

loading or off-axis loading of the sensor.  This is accomplished 

by applying a force to the sensor as close as possible to the 

center of the sensor.  In the event sensor is to be installed to 

measure a unit under test with a much larger area than that of 

the sensing surface of sensor, such as a large metal plate, it may 

be necessary to use an arrangement of two to four sensors in a 

measuring platform.  Independent sensor output can be 

monitored or the sensors can be connected electrically in 

parallel to measure the resulting summed forces when used in a 

multiple sensor type arrangement. 

 

TYPICAL ICP

 SYSTEM CONFIGURATION 

Sensors with built-in ICP

 circuitry require a constant-current 

excitation voltage for operation.  The enclosed specification 

sheet provides specific power requirements.  Required supply 

voltage is normally 20 to 30 VDC, while the constant current 

required ranges from 2 to 20 mA. 

 

PCB standard battery-powered signal conditioners are factory 

set at 2 mA and may be used to adequately drive a 5.5 kHz 

signal  using a typical PCB 29 pF/ft. cable to a desired 5-Volt 

full scale output, or about 175 feet (53 meters).  PCB line signal 

supplies are factory set at 4 mA (and adjustable from 2 to 20 

mA), enabling signals to be transmitted over hundreds of feet 

(meters), depending on the frequency of interest. 

 

It is necessary to supply the sensor with a 2 to 20 mA constant 

current at +20 to +30 VDC through a current-regulating diode 

or equivalent circuit, contained in all PCB signal conditioners.  

See Guide G-0001B for powering and signal conditioning 

information pertaining to all ICP

 instrumentation. 

 

Most of the line powered signal conditioners manufactured by 

PCB have an adjustable current feature allowing a choice of 

input currents from 2 to 20 mA.  In general, for lowest noise 

(best resolution), choose the lower current ranges.  When 

driving long cables (to several thousand feet(meters)), use 

higher current, up to 20 mA maximum.  Consult a factory 

applications engineer or local distributor to determine if higher 

current settings are required.   

 

For sensor operation, connect the sensor to the signal 

conditioner as shown in the typical ICP

 sensor systems below.  

Compete system operation requires the connection of the force 

sensor to a signal conditioner, then to a readout device 

(oscilloscope, meter, recorder, or A-to-D board) or to a readout 

device with built-in ICP

 sensor excitation.  Insert the cable 

center pin into the sensor electrical receptacle.  Tighten the 

coaxial cable to the sensor by turning the cable nut by hand to 

ensure good electrical contact.  

Do not spin the sensor onto 

the cable

 as this will fatigue the electrical center pin resulting 

in an intermittent connector or damaged cable.    

 

 

 

 

 

Figure 15 - Typical ICP

 System Configurations 

 

 

6.0   POLARITY 

 

Compressive forces upon an ICP

  force sensor produce a 

positive-going voltage output.  Tensile forces produce a 

negative-going voltage output.  Sensors with reversed polarity 

are available upon request. 

Содержание M221B02

Страница 1: ...comp 100 lb tension 50 mV lb Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...tallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description o...

Страница 7: ...ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Inst...

Страница 8: ...simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the spe...

Страница 9: ...eturning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for com...

Страница 10: ...t facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surface...

Страница 11: ...he left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total for...

Страница 12: ...itor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accom...

Страница 13: ...hundreds of feet meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent...

Страница 14: ...ant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents...

Страница 15: ...urned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specificat...

Страница 16: ...2 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side Mounting Thread 1 4 28 Female 1 4 28 Female All specifications are at room temperature unless otherwise specified In the inte...

Страница 17: ......

Отзывы: