background image

 

PCB Load & Torque, Inc.   

 Toll-Free in USA 866-684-7107   

 716-684-0001   

 www.pcbloadtorque.com

 

 

 

DUAL BRIDGE LOAD CELL OPERATION MANUAL 

 

2

 

1.0 INTRODUCTION 

 
Dual  bridge  fatigue-rated  load  cells  manufactured  by  PCB 
Load and Torque, Inc. are guaranteed to last 100 million fully 
reversed cycles (full tension through full compression). These 
load  cells  feature  high  accuracy  with  resistance  to  off-center 
loading, extraneous bending and side-loading forces. The load 
cells are manufactured using premium fatigue resistant steel or 
aluminum,  and  are  carefully  processed  to  ensure  mechanical 
and electrical integrity, as well as accuracy.  The nominal full 
scale output of the fatigue rated load cells is 2.0 mV/V. 
 
The  following  document  explains  how  to  properly  install  the 
dual  bridge  load  cell,  including  detailed  procedures  on 
specified  torques  installation,  adapter  thread  class  3  and 
tension pre-load specifications. 

 
 

2.0 SAFETY PRECAUTIONS 

 
Fatigue  rated  load  cells  are  designed  with  very  high  safety 
factors for overload, and of course, cyclic fatigue loads  when 
used  in  testing  programs.    However,  failure  of  the  load  cell 
structure  or  fasteners  used  in  the  load  cell  installation  may 
cause personal injury and equipment damage. 
 
It  is  important  to  provide  adequate  clearances  and  safety 
guards  or  shields  properly  surrounding  test  fixtures  where 
fatigue rated load cells are used.   
 
It is important to review the manufactured data when selecting 
rod ends and fasteners for load cell installation. Failure might 
occur due to combine measurement axis and extraneous loads 
if installation specifications are not followed.  All test fixtures 
designed  for  use  with  fatigue  rated  load  cells  should  be 
carefully evaluated for potential failure modes of the fixtures. 
 
It  is  important  to  use  threaded  fasteners  properly  when 
assembling test fixtures used with the load cells.  Fixtures with 
inadequately  designed  bolted  joints,  or  improperly  tightened 
threaded  fasteners  can  fail  before  the  parts  under  test  have 
completed the planned fatigue test cycles. 
 
 

3.0 OVERVIEW 

 
Dual  bridge  fatigue-rated  load  cells  are  specifically  designed 
for  repetitive  testing.  Applications  include  material  testing, 
component life cycle testing,  and structural testing.  The load 
cells  contain  an  extra  bridge  to  be  used  either  as  a  back-up 
incase  the  primary  bridge  fails  during  a  test,  or  just  for 
confirmation  of  the  primary  bridge  readings.  As  an  added 
benefit,  these  load  cells  are  extremely  resistant  to  extraneous 
bending and side loading forces 
 
 

3.1 Dimensions 

The following figure and table give the general outline 
dimensions of the dual bridge fatigue-rated load cells. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

11 

12 

 

15 

 

10 

R

13 

14 

Model Series 

1403 

1404 

1408 

1411 

Capacity (Lbf) 

5k 

12.5k, 25k 

50k 

100k 

Dimensions 

(see Figure 1) 

(in) 

(mm) 

(in) 

(mm) 

(in) 

(mm) 

(in) 

(mm) 

2.50 

63.5 

3.50 

88.9 

4.50 

114.3 

6.50 

165.1 

1.37 

34.8 

1.75 

44.5 

2.50 

63.5 

3.50 

88.9 

1.13 

28.7 

1.75 

44.5 

2.00 

50.8 

3.00 

76.2 

1.25 

31.8 

1.63 

41.4 

2.25 

57.2 

3.25 

82.6 

4.12 

104.6 

6.06 

153.9 

8.00 

203.2 

11.00 

279.4 

R

2.54 

64.6 

3.53 

89.7 

4.85* 

123.2* 

6.10* 

154.9* 

3.50 

88.9 

5.13 

130.2 

6.50 

165.1 

9.00 

228.6 

8

 

1.25 

31.8 

2.42 

61.4 

3.50 

88.9 

5.38 

136.5 

9

 

1.25 

31.8 

2.25 

57.2 

3.50 

88.9 

4.50 

114.3 

10 

45º 

30º 

22.5º 

22.5º 

11 

0.25 

6.4 

0.37 

9.4 

0.48 

12.2 

0.75 

19.1 

12 

0.03 

0.76 

0.03 

0.76 

0.03 

0.76 

0.03 

0.76 

13 

1/4-28x1 3/4 

screw 12 pt. 

Qty. 8 

3/8-24x2 1/4 

screw 12 pt. 

Qty. 12 

1/2-20x3 1/2 

screw 12 pt. 

Qty. 16 

3/4-16x4 1/2 

screw 12 pt. 

Qty. 16 

14 

5/8-18 

1 1/4-12 

1 3/4-12 

2 3/4-8 

Thread Depth 

0.88 

22.4 

1.37 

34.8 

2.12 

53.8 

2.56 

65.0 

15 

5/8-18 

1 1/4-12 

1 3/4-12 

2 3/4-8 

Thread Depth 

0.88 

22.4 

1.37 

34.8 

1.71 

43.4 

3.15 

80.0 

*Indicates dimension with connector protector. 

Figure 1

 

- Dimensions 

Table 2

 

– Dimension Values 

Dim. 

Description 

Total Height 

Load Cell Height 

Tension Base Height 

Load Cell Body 

Height 

Load Cell Diameter 

R

Max. Clearance 

Radius 

Bolt Circle Diameter 

Top Loading Surface 

Diameter 

Bottom Loading 

Surface Diameter 

10 

Angle Between 

Screws 

11 

Screw Head Height 

12 

Bottom Loading 

Surface Height 

13 

Screw for Tension 

Base 

14 

Load Cell Mounting 

Thread 

15 

Tension Base 

Mounting Thread 

 

Table 1

 

– Dimension 

Descriptions 

Содержание M1403-03ADB

Страница 1: ...D CELL DUAL BRIDGE METRIC Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 71...

Страница 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...LECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hy...

Страница 7: ...the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW Dual bridge fatigue rated load cells are specifically designed for repetitive testing Applications include material te...

Страница 8: ...ote that the performance of the load cell may be affected if the factory installed mounting base is removed or modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is...

Страница 9: ...ctor Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break o...

Страница 10: ...wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structur...

Страница 11: ...s are adjusted such that the error due to eccentric loading is less than 0 1 per inch of offset Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell...

Страница 12: ...tion certificate and adjust the span or gain until the display reads the force value stated on the certificate 4 Repeat steps 1 3 to verify that a valid calibration setting has been obtained 5 If poss...

Страница 13: ...ridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 11 2 1 Estimating Bridge B...

Страница 14: ...factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty...

Страница 15: ...in x 2 50 in 104 60 mm x 63 50 mm 8 Weight 6 4 lb 2 9 kg Mounting Thread No English Equivalent M16 x 2 0 Sensing Element Strain Gage Strain Gage Mounting Torque 200 in lb 23 Nm Deflection at Full Sca...

Страница 16: ...LOAD CELLS 5 8 18UNF 3B METRIC M LOAD CELLS M16X2 0 4H SPECIAL LOAD CELL 1 14UNS 3B 2 50 63 5 030 76 45 THREAD x 88 22 3 STANDARD LOAD CELLS 5 8 18 UNF 3B METRIC M LOAD CELLS M16 X 2 0 4H SPECIAL LOA...

Отзывы: