background image

 

Model Number

 

607A11

 

PLATINUM SWIVELER® INDUSTRIAL ICP® ACCELEROMETER

 

Revision: J 

ECN #: 41189

 

 

 

 
 
 

Performance 

ENGLISH 

SI 

Sensitivity(± 15 %)

 

100 mV/g

 

10.2 mV/(m/s²)

 

[2]

 

Measurement Range

 

± 50 g

 

± 490 m/s²

 

 

Frequency Range(± 3 dB)

 

30 to 600,000 cpm

 

0.5 to 10,000 Hz

 

 

Resonant Frequency

 

1500 kcpm

 

25 kHz

 

[1]

 

Broadband Resolution(1 to 10,000 Hz)

 

350 µg

 

3434 µm/sec

2

 

[1]

 

Non-Linearity

 

± 1 %

 

± 1 %

 

[3]

 

Transverse Sensitivity

 

 7 %

 

 7 %

 

 

Environmental 

Overload Limit(Shock)

 

5000 g pk

 

49,050 m/s² pk

 

 

Temperature Range

 

-65 to +250 °F

 

-54 to +121 °C

 

 

Enclosure Rating

 

IP68

 

IP68

 

 

Electrical 

Settling Time(within 1% of bias)

 

 2 sec

 

 2 sec

 

 

Discharge Time Constant

 

 0.3 sec

 

 0.3 sec

 

 

Excitation Voltage

 

18 to 28 VDC

 

18 to 28 VDC

 

 

Constant Current Excitation

 

2 to 20 mA

 

2 to 20 mA

 

 

Output Impedance

 

<150 Ohm

 

<150 Ohm

 

 

Output Bias Voltage

 

8 to 12 VDC

 

8 to 12 VDC

 

 

Spectral Noise(10 Hz)

 

8 µg/

Hz

 

78.5 (µm/sec

2

)/

Hz

 

[1]

 

Spectral Noise

(100 Hz)

 

5 µg/

Hz

 

49.1 (µm/sec

2

)/

Hz

 

[1]

 

Spectral Noise

(1 kHz)

 

4 µg/

Hz

 

39.2 (µm/sec

2

)/

Hz

 

[1]

 

Electrical Isolation(Case)

 

>10

8

 Ohm

 

>10

8

 Ohm

 

 

Physical 

Size (Hex x Height)

 

9/16 in x 0.97 in

 

14 mm x 24.6 mm

 

 

Weight(without cable)

 

1.1 oz

 

31 gm

 

[4]

 

Mounting

 

Stud

 

Stud

 

 

Mounting Thread

 

1/4-28 Male

 

1/4-28 Male

 

[5]

 

Mounting Torque(stud)

 

3 to 4 ft-lb

 

4.1 to 5.4 Nm

 

[6][7]

 

Mounting Torque

(hex nut)

 

2 to 3 ft-lb

 

2.7 to 4.1 Nm

 

 

Sensing Element

 

Ceramic

 

Ceramic

 

 

Sensing Geometry

 

Shear

 

Shear

 

 

Housing Material

 

Stainless Steel

 

Stainless Steel

 

 

Sealing

 

Welded Hermetic

 

Welded Hermetic

 

 

Electrical Connector

 

Molded Integral Cable

 

Molded Integral Cable

 

 

Electrical Connection Position

 

Side

 

Side

 

 

Cable Length

 

10 ft

 

3.0 m

 

 

Cable Type

 

Polyurethane

 

Polyurethane

 

[8]

 

 

[9]

 

All specifications are at room temperature unless otherwise specified. 
In the interest of constant product improvement, we reserve the right to change specifications without notice. 

ICP

®

 is a registered trademark of PCB Group, Inc. 

 

OPTIONAL VERSIONS 

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

 

 

 
EX 

- Hazardous Area Approval- contact factory for specific approvals

 

Hazardous Area Approval

 

Exia IIC T4, AExia IIC, T4

 

Exia IIC T4, AExia IIC, T4

 

Hazardous Area Approval

 

EEx nL IIC T4, -40°

C

Ta

121°C, II 1 G

 

EEx nL IIC T4, -40°

C

Ta

121°C, II 1 G

 

Hazardous Area Approval

 

Cl I, Div I, Groups A, B, C, D; 

Cl II, Div I, Groups E, F, G; Cl 

III, Div I

 

Cl I, Div I, Groups A, B, C, D; 

Cl II, Div I, Groups E, F, G; Cl 

III, Div I

 

Hazardous Area Approval

 

Cl I, Div 2, Groups A, B, C, D; 

ExnL IIC T4, AExnA IIC T4

 

Cl I, Div 2, Groups A, B, C, D; 

ExnL IIC T4, AExnA IIC T4

 

Hazardous Area Approval

 

EEx nL IIC T4, -40°

C

Ta

121°C, II 3 G

 

EEx nL IIC T4, -40°

C

Ta

121°C, II 3 G

 

 
M 

- Metric Mount

 

Supplied Accessory : Model M080A159A Mounting stud, 1/2-20 to M6 x 1 (1) 

 
TO 

- Temperature Output

 

Temperature Output Range

 

+36 to +250 °F

 

+2 to +121 °C

 

Temperature Scale Factor

 

5.56 mV/°F + 32

 

+10 mV/°C

 

Electrical Connector

 

Molded Integral Cable

 

Molded Integral Cable

 

Electrical Connections(Red)

 

Acceleration Output

 

Acceleration Output

 

Electrical Connections

(Black)

 

Ground

 

Ground

 

Electrical Connections

(White)

Temperature Output

 

Temperature Output

 

 

NOTES:

[1] Typical.
[2] Conversion Factor 1g = 9.81 m/s².
[3] Zero-based, least-squares, straight line method.
[4] Measured with mounting stud.
[5] 1/4-28 has no equivalent in S.I. units.
[6] 1/8" hex Allen key required for English version, 3mm hex Allen key required for metric 

version.

[7] Stud torque must exceed sensor hex nut torque to ensure proper dismantling.
[8] Twisted shielded pair.
[9] See PCB Declaration of Conformance PS023 or PS060 for details.

 

 

 

SUPPLIED ACCESSORIES:  

Model 080A156 Mounting Base (1) 
Model ICS-2 NIST-traceable single-axis single-point amplitude response calibration at 6000 
cpm (100 Hz) (1) 

Entered: AP

 

Engineer: jg

 

Sales: EGY

 

Approved: BAM

 

Spec Number:

 

Date: 4/11/2013

 

Date: 4/11/2013

 

Date: 4/11/2013

 

Date: 4/11/2013

 

12650 

 

3425 Walden Avenue, Depew, NY 14043

Phone: 800-959-4464 
Fax: 716-684-3823 
E-Mail: [email protected]

Содержание IMI SENSORS TOM607A11

Страница 1: ...ICP Accelerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 959 4464 24 hour SensorLine 716 684 0001 Fax 716 684 38...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...ON A ECN NUMBER 49766 Operating Guide with Enclosed Warranty Information 3424 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 959 4IMI Piezoelectric IC...

Страница 7: ...ply low frequency and high frequency measurements Shear design sensors are preferred because of their inherent insensitivity to adverse environmental influences such as case or base strain and thermal...

Страница 8: ...wer gain which is extremely close to unity and independent of input voltage Also the diode can be changed to supply higher currents for driving long cable lengths Constant current diodes as shown in F...

Страница 9: ...nics However when operating AC coupled signal conditioners with readout devices having an input impedance of less than one megaohm thelow frequencyrangemaybeaffected 4 1 STANDARD STUD MOUNT This mount...

Страница 10: ...derside of the mounting base Firmly press down on the assembly to displace any extra adhesive remaining under the base See Figure7 METHOD 2 Direct Adhesive Mount For restrictions of space or for conve...

Страница 11: ...interference or radio frequency interference EMI or RFI To avoid ground loops thereshouldonlybeonegroundin the system It is recommended for permanent installations that the sensor becase isolatedand...

Страница 12: ...ate Also any current not used by the cable goes directly to power the internal electronicsandwillcreateheat Thismaycausethesensorto exceed its maximum temperature specification For this reason donotsu...

Страница 13: ...er the meter or LED and allow the sensor to power up If a faulty condition is monitored first check all system connections then check the functionality of the cable and signal conditioner If the syste...

Страница 14: ...and former MIL STD 45662A and uses equipment directly traceable to NIST This assures an accurate calibration of relevant specifications The following includes a broad overview of the Back to Back Cal...

Страница 15: ...A of constant current to extend battery life while continuous monitoring systems offer more current in order to drive longer cables Figure 16 shows a typical system for a sensor with integral electron...

Страница 16: ...eive assistance via e mail at imi pcb com or visit our website at www pcb com 11 0 CUSTOMER SERVICE WARRANTY IMI a division of PCB Piezotronics guarantees Total Customer Satisfaction If at any time fo...

Страница 17: ......

Страница 18: ...ersions have identical specifications and accessories as listed for the standard model except where noted below More than one option may be used EX Hazardous Area Approval contact factory for specific...

Страница 19: ...ALL MOUNTING SURFACES SEE ARROWS FIG 1 2 FIG 2 TIGHTEN THE MOUNTING STUD USING THE ALLEN KEY TORQUE THE MOUNTING STUD TO WITHIN 3 TO 4 FT LBS 4 1 TO 5 4 Nm FIG 3 THREAD THE SENSOR S HEX NUT ONTO THE...

Страница 20: ...ANCES IN METRIC 1 LENGTH 1 1 0 1 LENGTH 5 2 0 5 LENGTH 100 6 0 100 LENGTH 1 0 2 54cm LENGTH 30 5cm 2 54cm 0 30 5cm LENGTH 1 5m 5 1cm 0 1 5m LENGTH 30 5m 15 2cm 0 30 5m LENGTH 30 5cm 0 SWIVELERTM ACCEL...

Отзывы: