Drawing Number: 21102 

Revision: B 

 

 

 

 

OPERATION MANUAL FOR 

2  

Series 138 

 

 

Some models of the Series 138 feature a tygon tube 

closed at one end with conical-shaped delrin plug. The 

plug features a small hole for attachment of a weight. 

Other models feature a one-piece molded closed tube. 

 

2.0

 

INSTALLATION, CABLE 

ATTACHMENT, AND SEALING 

THE CABLE/CONNECTOR 

ASSEMBLY 

Various methods are used to mount the 138 sensors for 

use in liquids.  Care should be taken to prevent 

conductive  liquids  from  entering  the  cable  or 

connector  assemblies,  where  they  could  cause 

corrosion,  short-circuiting  or  degradation  of 

insulation resistance. 

 

For low-pressure (<500psi) dynamic measurements, a 

standard  low-noise  coaxial  cable  with  an  extruded 

teflon jacket and a 10-32 connector can  be  used. Heat 

shrink tubing, RTV silicon rubber,  or sometimes 

silicone  grease  can  make  an  adequate  seal  in  an 

environment with low static pressure. 

 

For high static pressures due to deep submersion or use 

in a pressure vessel, alternate methods of sealing the 

cable/connector may be required.  A method of 

waterproofing the cable/connector assembly of a 138 

which also provides protection to the cable is to put a 

flexible  tube  (plastic,  tygon,  etc.)  around  the  cable 

and cable/connector assembly.  Tie the tube at the 

metal housing with monofilament or fine steel wire. 

The opposite end of the tube must be above the level 

of the liquid or sealed around the cable to prevent 

liquid from entering the protective tubing. Filling the 

protective tube with a non-conducting liquid such as 

Dow Coming silicone oil can be done to prevent the 

tube from collapsing. This will also equalize pressure 

inside and outside the tubing, helping to prevent a 

conductive liquid (such as water) from entering. 

 

In some blast applications, when using a standard 10- 

32 micro cable/connector, the high shock can flex the 

cable/connector joint enough to move the center pin of 

the 10-32 cable connector.   This can cause noise on 

the output signal.  Protecting the connection from 

shock may be required. 

 

 

 

 

 

 

 

 

PCB Model 070A03 adaptor can be used to adapt the 

10-32 micro jack on the 138 to a BNC jack.  A  heavier 

coaxial cable (eg. RG-58/U) can now be used  in place 

of the micro cable.   Because of the large mass of the 

070A03 compared to the 10-32 jack, this connection 

should  be  sealed  and  protected  from  shock and 

turbulence. Putting a two to three inch diameter of duct 

seal,  wrapped  with  a  waterproof tape,  will  seal  the 

connection  and  provide  shock  protection  at  the 

cable/connector/ amplifier assembly. 

 

PCB Model 070A09 Solder Connector Adaptor can 

also be used to attach a short length (~two to three 

inches) of ribbon cable to a 138.   The ribbon cable can 

then  be  soldered  to  an  extension  cable.  The   ribbon 

cable and both solder connections should be  sealed 

with a suitable waterproofing  material. Various types 

of sealing materials such as polyethylene and flexane 

have  been  shown  to  be  effective  for  sealing  cables, 

connectors, etc. 

 

2.1

 

MOUNTING OR  SUSPENDING 

THE  SENSOR 

After  the  cable/connector  assembly  is  sealed,  the 

sensor  can  be  suspended  in  the  liquid  where  the 

measurement  is  to  take  place.  For  low  dynamic 

pressure measurements (less than 500 psi), models 

featuring a small hole in the conical end piece of the 

tube allow attachment of a light weight (~two lb.) to 

suspend the sensor in a vertical position. 

 

The line should be  of light material (monofilament, 15 

lb. test) to enable it to break easily.  This way, the 

sensor does not pull apart in the turbulence caused by 

the collapse of gas bubbles resulting from a shock 

wave. 

 

For higher pressure dynamic measurements, alternate 

methods of suspending the sensor should be used to 

minimize  damage  from  turbulence.  Taping  or  

attaching (with cable ties) the sensor amplifier  housing 

to a braided steel wire is a method that is suited for use 

in higher dynamic pressure measurements. 

Содержание ICP W138A02

Страница 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail i...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...ducted over a single conductor coaxial cable with a shield serving as signal return Special low noise cables are not required Series 138 is used by the military for underwater explosive testing commer...

Страница 7: ...high shock can flex the cable connector joint enough to move the center pin of the 10 32 cable connector This can cause noise on the output signal Protecting the connection from shock may be required...

Страница 8: ...itioner If indicator is in green section of indicator panel the IC amplifier is producing proper bias 8 to 14 VDC cable connections are normal and the system is ready to operate If the needle moves to...

Страница 9: ...sor l Keep connections clean and dry 2 Do not exceed specified maximum range 3 Do not subject sensor to temperaturesexceeding 100 F 4 Do not apply voltage to sensor withoutcurrent limiting diodes or o...

Страница 10: ...nt Tourmaline Tourmaline HousingMaterial Stainless Steel Stainless Steel Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 61 oz 17 0 gm 4 All specifications are at room temperature...

Страница 11: ......

Отзывы: