Drawing Number: 40955 
Revision: NR 
ECO#: 39372 

 

OPERATION MANUAL FOR 

FREE FIELD BLAST PRESSURE PROBE 

Models 137B21X, B22X, B23X, B24X 

 

 2 

Switch  power  on  and  observe  reading  of  bias 
monitoring  voltmeter  on  front  panel  of  signal 
conditioner. 
 
If indicator is in green section of indicator panel, the 
IC amplifier is producing proper bias (8 to 14 VDC), 
the  cable  connections  are  normal,  and  the  system  is 
ready to operate. 
 
If  the  pointer  moves  into  the  red  area  of  the  fault 
monitor meter, output is zero and a short is indicated.  
The  short  could  be  located  in  amplifier  cable 
connectors or power unit. 
 
If the pointer moves into the yellow area of the fault 
monitor  meter,  an  open  circuit  is  indicated  with  full 
power supply voltage. 
 
An  open  circuit  could  be  the  result  of  a  faulty 
amplifier,  an  open  cable  or  open  connectors.    Check 
to be sure connectors are properly mated.  Some PCB 
power  units  feature  fault  LEDs  in  place  of  a  fault 
meter.    Instead  of  a  pointer,  the  fault  LED  will  light 
when an open or short circuit condition is present. 
 
Allow the sensor to thermally stabilize for about one 
minute.    A  signal  drift  may  occur  when  the  cable  is 
connected  to  the  readout  instrument.    This  drift 
occurs  during  charging  of  the  coupling  capacitor  in 
the  power  unit.    The  signal  will  stabilize  in  several 
minutes.  Proceed with measurements. 
 
Most  line-powered  signal  conditioners  manufactured 
by PCB have an adjustable current feature allowing a 
choice of input currents from 2 to 20 mA.  In general, 
for  the  lowest  noise  (best  resolution),  choose  the 
lower current ranges.   
 
In order to maintain a high frequency response of the 
system  when  driving  long  cables  (to  hundreds  or 
thousands of feet), use a higher current, up to 20 mA 
maximum.    Use  of  low-capacitance  cable  (i.e.  RG 
62/U)  and  impedance-matching  variable  resistors  are 
suggested  to  improve  long  cable  performance 
characteristics.    See  Guide  G-0001  for  a  discussion 
on cable considerations. 
 

3.1 

POLARITY 

 
The  137B2XX  Series  produces  a  positive-going 
output voltage for increasing pressure output. 
 
 

 

4.0 

LOW-FREQUENCY RESPONSE 

 
Consult  Section  7.0  in  General  Guide  G-0001  for 
detailed  explanation  of  low-frequency  characteristics 
of ICP

®

 instrumentation. 

 

5.0 

CALIBRATION 

 
Do not try to calibrate the Series 137B2XX statically.  
The  Series  137B2XX  is  calibrated  dynamically  by 
PCB using a hydraulic pulse technique. 
 
For  best  accuracy,  use  the  calibration  certificate 
supplied.    Factory  recalibration  is  available  for  a 
nominal charge. 
 

6.0 

MAINTENANCE 

 
Although  ICP

®

  sensors  have  low-output  impedance 

and are not usually affected by moisture, in extremely 
damp  environments  it  may  be  well  to  protect  cable 
connections with shrink tubing. 
 
It  is  well  to  observe  the  following  precautions  in 
using the sensor: 
 
1. 

Do not exceed specified maximum range. 

 
2. 

Do  not  subject  sensor  to  temperatures 

exceeding temperature shown on specification sheet. 
 
3. 

Do  not  apply  voltage  to  sensor  without 

current-limiting diodes or other current protection. 
 
4. 

Do not apply more than 20 mA of current to 

the sensor. 
 
Field  repair  of  the  piezoelectric  element  or  amplifier 
of  the  Series  137B2XX  is  not  practicable.    Thus, 
should  a  sensor  in  this  series  require  servicing,  refer 
to the warranty sheet. 
 
CAUTIONARY  NOTE:    If  sensors  are  left  outside 
overnight, they should be coated with silicone oil and 
covered.  The protective silicone oil or grease coating 
should also be applied when the sensors are operated 
in a humid or rainy environment. 
 
 
 
 
 