background image

 

Drawing Number: 40955 
Revision: NR 
ECO#: 39372 

 

OPERATION MANUAL FOR 

FREE FIELD BLAST PRESSURE PROBE 

Models 137B21X, B22X, B23X, B24X 

 

 1 

1.0 

INTRODUCTION 

 
The Series 137B2XX Free Field Blast ICP

®  

Pressure 

Probe is designed for measuring free field blast and 
shock tunnel pressure time profiles. 
 
Designed  with  an  ICP

®

  built-in  source  follower 

amplifier,  the  137B2XX  Series  offers  sensitivities  of 
1, 20, 10 and 100 mV/psi with full-scale ranges from 
0 to 50, 250, 500 or 5000 psi.  The 137B2XX Series 
is  constructed  with  a  stable  quartz  piezoelectric 
element in an Invar housing 
 

2.0 

INSTALLATION 

 
In  field  blast  measurements,  mount  the  Series 
137B2XX pressure probe in an axial direction to the 
blast  source.    The  probes  should  be  elevated  high 
enough  off  of  the  ground  such  that  ground  does  not 
interfere with the measured pressure. See photo 1. 
 

 

Photo 1 

 

The  sensing  surface  should  be  aligned  in  a  vertical 
plane with respect to the ground as shown in photo 2.  
Avoiding  direct  exposure  of  the  diaphragm  to 
sunlight  is  good  practice  so  that  changing 
temperatures throughout a test day do not change the 
sensitivity  of  the  sensor.  In  any  case  care  must  be 
taken such that the sensor’s diaphragm is not oriented 
down to avoid ground reflections. 
 
In some cases, where flash temperatures such as those 
generated  by  blasts  and  shock  fronts  are  present,  it 
may be necessary to thermally insulate the diaphragm 
to minimize signals generated by these effects. 
 

 

Photo 2 

 
Common black vinyl electrical tape has been found to 
be  an  effective  thermal  insulating  material  in  many 
cases.    One  layer  may  be  used  across  the  end  of 
diaphragm and adaptor.  See photo 3. 
 

 

Photo 3 

 
A  silicone  rubber  coating  approximately  0.010  inch 
(0.25  mm)  thick  has  also  been  proven  effective  in 
many applications.  General Electric RTV type 106 is 
recommended  (PCB  Model  065A67).    Apply  the 
rubber  coating  to  the  surface  of  the  diaphragm  and 
allow it to cure in accordance with the manufacturer’s 
instructions. 
 

3.0 

OPERATION 

 
Using  suitable  cable,  connect  the  sensor  to  one  of 
PCB's  Series  480,  481,  482,  483,  or  484  signal 
conditioners  or  equivalent  power  supply.    If  using  a 
PCB  line  power  unit,  the  equipment  will  supply  the 
137B2XX  with  an  adjustable  2  to  20  mA  constant 
current  at  +22  to  +30  VDC  through  a  current-
regulating  circuit.    If  using  a  battery-operated  unit, 
the constant current is 2 mA.  See Guide G-0001 for 
powering  and  signal  utilization  information  pertinent 
to all ICP

®

 instrumentation. 

 

 