background image

CODE

IDENT. NO.

52681

DWG. NO.

SCALE:

SHEET

DRAWN

CHECKED

ENGINEER

TITLE

UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:

DIMENSIONS IN MILLIMETERS

[ IN BRACKETS ]

XX

    

±

 

0.13

ANGLES

  

 

2 DEGREES

3425 WALDEN AVE. DEPEW, NY 14043

(716)  684-0001 E-MAIL:  [email protected]

DIMENSIONS  IN INCHES

DECIMALS     XX

  

 

±.01

ANGLES

  

 

2 DEGREES

FILLETS AND RADII 

 .003 - .005 

DECIMALS     X

   

 ±

 

0.3

FILLETS AND RADII 

0.07 - 0.13 

INSTALLATION DRAWING

65031

1 OF 1

2X

MODEL EX357E92

ACCELEROMETER

XXX   ±.005

JES

9/5/19

JDM

9/5/19

YHK

9/5/19

DIRECTION OF 

ACCELERATION

TO PROVIDE A 

NEGATIVE

OUTPUT SIGNAL

1

1

2

2

3

3

4

4

A

A

B

B

65

03

1

PCB Piezotronics Inc. claims proprietary rights in

the information disclosed hereon.  Neither it nor any

reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.

REVISIONS

REV

DESCRIPTION

DIN

A

EX357E92 WAS 357E92

49874

.56 [14.2]

.28 [7.1]

.95 [24.1]

.23 [5.7]

120.0

4.0 3048

102

[

]

.72 [18.2]

1.28 [32.5]

.25 [6.4]

Ø

.196 THRU



.396 

.180

4.98 THRU



10.06 

4.57

[

]

MINERAL OXIDE INSULATED
TRIAXIAL HARDLINE CABLE

5.)    CABLE MINIMUM BEND RADIUS: .25 [6.4]
 
4.)    RECOMMENDED MOUNTING TORQUE: 10-20 INCH POUNDS [113-225 Ncm]

 
3.)    FOR BEST RESULTS, PLACE A THIN LAYER OF SILICONE GREASE (OR EQUIVALENT) ON INTERFACE PRIOR TO MOUNTING
 
        MOUNTING SURFACES SHOULD BE FLAT TO WITHIN .001[.03] TIR ACROSS A 2" DIAMETER AREA WITH A 
        SURFACE FINISH OF 63 [1.61] OR BETTER FOR BEST RESULTS

 
        DRILL PERPENDICULAR TO MOUNTING SURFACE TO WITHIN ± 1°

1

2

MOUNTING HOLE PREPARATION:

Ø.159[Ø4.04] 

(X) +.130 [3.30]

10-32 UNJF - 3B 

(X) +.035 [.89]

WHERE: (X)=DISTANCE MOUNTING SCREW

EXTENDS PAST MOUNTING SURFACE

[I.E. SCREW INSERTION DEPTH]

2

1

10-32 COAXIAL CONNECTOR

Содержание EX357E92

Страница 1: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 2: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 3: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 4: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 5: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Страница 6: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Страница 7: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Страница 8: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Страница 9: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Страница 10: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Страница 11: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Страница 12: ...Hermetic Welded Hermetic Size Height x Length x Width 56 in x 95 in x 56 in 14 2 mm x 24 1 mm x 14 2 mm Weight without cable 1 6 oz 45 gm 1 Electrical Connector Integral Hardline Cable Integral Hardli...

Страница 13: ...eof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN A EX357E92 WAS 357E92 49874 56 14 2 28 7 1 95 24 1 23 5 7 120 0 4 0 3048 102 72 18 2 1...

Отзывы: