3

 

When  an  ICP

®

  transducer  (or  in-line  amplifier)  is  connected  to  the  INPUT  jack,  the  front  monitor  meter  indicates 

approximately  midscale  (green)  if  the  transducer  or  amplifier  is  functioning  properly.  It  does  not  read  midscale  if  the 
transducer bias is 3 to 5 volts. 
 
If  a  cable  is  faulty  (open),  or  the  sensor's  built-in  amplifier  is  open,  the  meter  indicates  a  yellow  (full  scale)  reading. 
When a cable or transducer is shorted, the meter indicates zero volts (red). 
 
Immediately after connecting a readout instrument (oscilloscope, meter, or recorder) to the OUTPUT jack, the coupling 
capacitor  begins  charging  through  the  input  resistance  of  the  readout  instrument.  This  charging  causes  an  apparent 
"drifting" of the output signal until the capacitor is fully charged. The normal amount of leakage through the coupling 
capacitor usually results in a <20 mV maximum DC offset.  
 

4.0 Coupling Time Constant, AC Coupled 

 

The coupling time constant (TC) is the product of the coupling capacitor (47 µF) and the input resistance of the readout 
instrument, in parallel with the 274 kilohm shunt resistor. Typically, this is ten seconds, assuming the input resistance of 
the readout measurement is one megohm. 
 
In most cases, it is desirable to keep the coupling TC long, with respect to the transducer discharge TC, to minimize the 
effect of the coupling TC on low frequency response.  
 

4.1 Driving Long Cables 

 
When  driving  long  cables,  it  may  be  necessary  to  increase  the  constant  current  drive  to  the  transducers.  482A21  is 
normally  supplied  with  the  constant  current  output  to  the  transducer  set  at  4  mA  nominal.  This  is  adequate  for  most 
laboratory and field applications. Special situations, such as driving long cables (beyond 100 ft) with high frequency or 
fast rise time pulses, may require increasing the transducer drive current up to 20 mA. 
 
When driving fast rise time pulses over long lines, system performance may be optimized by "tuning" the drive current 
to  the  line;  i.e.,  by  finding  the  best  current  setting  for  the  particular  set  of  physical  parameters  established  by  the 
transducer, line length, line termination, pulse rise time, etc. 
 
The optimum current setting is best determined by experimentation with your particular test setup. A good rule of thumb 
is to use the lowest current consistent with satisfactory results to minimize transducer self-heating and noise. 
 
Another  technique  requires  the  use  of  impedance  matching  resistor  model  073A  connected  immediately  after  the 
transducer. This variable resistor is then used to "tune" the current to the line.  
 
One of the best approaches is to test the long line with a signal generator set at a frequency of interest. 
 

4.2 Setting the Constant Current 

 
To set the constant current, locate the current adjust potentiometer on the rear panel. Connect a 0 to 30 mA DC ammeter 
(or multimeter) to the INPUT jack (as shown in Figure 3). The constant current value is read directly on the ammeter. 
Using a small regular screwdriver, vary the setting of the current adjust pot to set the current at a new level. 

 
CAUTION:   

It  may  be  possible  to  exceed  20  mA  slightly.  Do  not  do  so;  to  operate  an  ICP

®

  transducer  or  amplifier 

above 20 mA may damage the equipment. 
 

Содержание 482A21

Страница 1: ...signal cond unity gain BNC Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 7...

Страница 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Страница 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...grounded person The solution for product safety 1 Connect the cables only with the AC power off 2 Temporarily short the end of the cable before attaching it to any signal input or output Caution 2 ESD...

Страница 7: ...ought out to the OUTPUT jack The bias monitor located on the front panel consists of a color coded panel meter that permanently monitors the sensor bias level See sensor specification for correct sens...

Страница 8: ...tant current drive to the transducers 482A21 is normally supplied with the constant current output to the transducer set at 4 mA nominal This is adequate for most laboratory and field applications Spe...

Страница 9: ...ative concerning the difficulty to ascertain if an on site procedure rectifies the problem If repair is indicated contact PCB to request a Return Materials Authorization RMA number from the factory An...

Страница 10: ...select matches the channel of interest Output signal seems to drift Make sure the input impedance of the readout device scope or data acquisition system is approximately 1 M Low frequency response not...