background image

 

MANUAL #: 19179 
MANUAL REV: B 
ECO: 43785 

 

MODEL 482A21 LINE POWERED SIGNAL CONDITIONER

 

5

 

482A21 and 482A22 Troubleshooting Guide 

Problem 

Test/Solution 

Unit  does  not  appear  to  be  powered 
up/ No meter movement. 

- Power switch in the on position. 
-  Test  ‘sensor’  excitation  current.  Measure  the  current  on  the  ‘sensor’  input 
BNC  from  the  center  conductor  to  the  shell.  Adjust  ‘Iadj’  (rear  panel)  with 
screwdriver  turning  both  clockwise  and  counterclockwise.  Reference  section 
‘Setting the Constant Current.’ 
- Make sure the bias monitor channel select matches the channel of interest. 

Output signal seems to drift. 

-  Make  sure  the  input  impedance  of  the  readout  device,  scope  or  data 
acquisition system is approximately 1 MΩ. 

Low  frequency  response  not  meeting 
specification 

or 

amplitude 

is 

attenuated. 

- The input impedance of the readout device may be too low. May need to use a 
buffered signal conditioner, for example 482C16. 

Meter  does  not  seem  to  show  sensor 
bias voltage. 

- Match the channel select with channel of interest. 
- Measure the sensor bias while the sensor is attached to the unit. This requires 
a  ‘T’  connector  to  allow  both  the  sensor  and  the  measuring  device  to  be 
attached. If problem persists, return for repair. 

 
MANUAL NUMBER: 19179 
MANUAL REVISION:  
ECO NUMBER:  

 

Содержание 482A21

Страница 1: ...signal cond unity gain BNC Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 7...

Страница 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Страница 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...grounded person The solution for product safety 1 Connect the cables only with the AC power off 2 Temporarily short the end of the cable before attaching it to any signal input or output Caution 2 ESD...

Страница 7: ...ought out to the OUTPUT jack The bias monitor located on the front panel consists of a color coded panel meter that permanently monitors the sensor bias level See sensor specification for correct sens...

Страница 8: ...tant current drive to the transducers 482A21 is normally supplied with the constant current output to the transducer set at 4 mA nominal This is adequate for most laboratory and field applications Spe...

Страница 9: ...ative concerning the difficulty to ascertain if an on site procedure rectifies the problem If repair is indicated contact PCB to request a Return Materials Authorization RMA number from the factory An...

Страница 10: ...select matches the channel of interest Output signal seems to drift Make sure the input impedance of the readout device scope or data acquisition system is approximately 1 M Low frequency response not...

Страница 11: ...al Connector DC Power Input DIN Jack DIN Jack Size Height x Width x Length 6 3 in x 2 4 in x 11 in 16 cm x 6 1 cm x 28 cm Weight 1 51 lb 685 gm All specifications are at room temperature unless otherw...

Страница 12: ......

Страница 13: ......

Отзывы: