background image

 

Model Number

 

378B20

 

ICP MICROPHONE SYSTEM

 

Revision: A 

ECN #: 42896

 

 

 

[7]

 

Performance 

ENGLISH 

SI 

Nominal Microphone Diameter

 

1/2"

 

1/2"

 

 

Frequency Response Characteristic

 

Random Incidence

 

Random Incidence

 

 

Open Circuit Sensitivity

 

50 mV/Pa

 

50 mV/Pa

 

[3]

 

Open Circuit Sensitivity

(+/-1.5 dB)

 

-26 dB re 1 V/Pa

 

-26 dB re 1 V/Pa

 

[3]

 

Frequency Range(+/-1 dB)

 

7 to 6300 Hz

 

7 to 6300 Hz

 

 

Frequency Range

(+/-2 dB)

 

3.75 to 12,500 Hz

 

3.75 to 12,500 Hz

 

 

Lower Limiting Frequency

 

1.0 to 3.0 Hz

 

1.0 to 3.0

 

 

Inherent Noise(Linear)

 

<20 dB re 20 µPa

 

<20 dB re 20 µPa

 

 

Inherent Noise

 

<16.5 dB(A) re 20 

µPa

 

<16.5 dB(A) re 20 

µPa

 

 

Dynamic Range(3% Distortion Limit)

 

>135 dB re 20 µPa

 

>135 dB re 20 µPa

 

 

Dynamic Range

(Maximum without Clipping)

 

138 dB re 20 µPa

 

138 dB re 20 µPa

 

[2][4]

TEDS Compliant

 

Yes

 

Yes

 

[5]

 

Environmental 

Temperature Range(Operating)

 

-40 to +176 °F

 

-40 to +80 °C

 

 

Temperature Coefficient of Sensi14 to +158°F (-10 to +70°C))

0.005 dB/°F

 

0.009 dB/°C

 

[2][3]

Static Pressure Coefficient

 

-0.01 dB/kPa

 

-0.01 dB/kPa

 

[2][3]

Humidity Coefficient of Sensitivity(0 to 100%, non-condensing)

 

± 0.001 dB/%RH

 

± 0.001 dB/%RH

 

[3]

 

Influence of Axial Vibration(0.1g (1 m/s²))

 

63 dB re 20 µPa

 

63 dB re 20 µPa

 

[2]

 

Electrical 

Polarization Voltage

 

0 V

 

0 V

 

[1]

 

Excitation Voltage

 

20 to 30 VDC

 

20 to 30 VDC

 

 

Constant Current Excitation

 

2 to 20 mA

 

2 to 20 mA

 

 

Output Bias Voltage

 

10 to 14 VDC

 

10 to 14 VDC

 

 

Output Impedance

 

<50 Ohm

 

<50 Ohm

 

 

Maximum Output Voltage

 

+/-7 Vpk

 

+/-7 Vpk

 

 

Physical 

Housing Material

 

Stainless Alloy

 

Stainless Alloy

 

 

Venting

 

Rear

 

Rear

 

[6]

 

Electrical Connector

 

BNC Jack

 

BNC Jack

 

 

Mounting Thread(Grid)

 

0.5 - 60 UNS

 

0.5 - 60 UNS

 

 

Size (Diameter x Height)(with grid)

 

0.52 in x 3.62 in

 

13.2 mm x 91.9 

mm

 

 

Size (Diameter x Height)

(without grid)

 

0.50 in x 3.58 in

 

12.7 mm x 90.9 

mm

 

 

Weight

 

1.63 oz

 

45.8 gm

 

[2]

 

All specifications are at room temperature unless otherwise specified. 
In the interest of constant product improvement, we reserve the right to change specifications without notice. 

 

OPTIONAL VERSIONS 

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

 

 

 
TLD 

- TEDS Capable of Digital Memory and Communication Compliant with IEEE 1451.4

 

Optional Accessory : Model ACS-63 Calibration (with TEDS) of Precision Condenser 
Microphones and Preamplifiers together (mated pair). (1) replaces Model ACS-42 

 

 

NOTES:

[1] Prepolarized
[2] Typical.
[3] re 250 Hz
[4] Peak.
[5] TEDS Capable Digital Memory and Communication, compliant with IEEE P1451.4
[6] Venting through Preamp.
[7] See PCB Declaration of Conformance PS064 for details.

 

 

 

SUPPLIED ACCESSORIES:  

Model ACS-42 Calibration of microphone with preamplifer (1) 

Entered: AP

 

Engineer: MJN

 

Sales: MV

 

Approved: MT

 

Spec Number:

 

Date: 5/7/2014

 

Date: 5/7/2014

 

Date: 5/7/2014

 

Date: 5/7/2014

 

57826 

 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001 
Fax: 716-684-0987 
E-Mail: [email protected]

Содержание 378B20

Страница 1: ...e 377B20 426E01 installation and operating manuals that comprise a Model 378B20 ICP Microphone System kit For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 8...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Страница 5: ...ponse reactions of the human ear to the pressure fluctuations Some examples of typical sound pressure levels are shown in the table below dB Pressure Example 0 0 00002 Pa Threshold of Hearing 60 0 02...

Страница 6: ...and noise cones the fre quency response of the microphone will be attenuated at higher frequencies Clamps and stands When holding a microphone in place it is important to minimize the influence of the...

Страница 7: ...2LA manufacturers Our internal metrology laboratory is certified to ISO 9001 accredited by A2LA to AN SI IEC 17025 and ANSI NCSL Z540 1 complies with ISO 10012 1 and former MIL STD 45662A and uses equ...

Страница 8: ...VDC Output Impedance 50 Ohm 50 Ohm Maximum Output Voltage 7 Vpk 7 Vpk Physical Housing Material Stainless Alloy Stainless Alloy Venting Rear Rear 6 Electrical Connector BNC Jack BNC Jack Mounting Thre...

Страница 9: ......

Страница 10: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E m...

Страница 11: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 12: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 13: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Страница 14: ...ponse reactions of the human ear to the pressure fluctuations Some examples of typical sound pressure levels are shown in the table below dB Pressure Example 0 0 00002 Pa Threshold of Hearing 60 0 02...

Страница 15: ...and noise cones the fre quency response of the microphone will be attenuated at higher frequencies Clamps and stands When holding a microphone in place it is important to minimize the influence of the...

Страница 16: ...2LA manufacturers Our internal metrology laboratory is certified to ISO 9001 accredited by A2LA to AN SI IEC 17025 and ANSI NCSL Z540 1 complies with ISO 10012 1 and former MIL STD 45662A and uses equ...

Страница 17: ...20 Pa 1 Electrical Capacitance Polarized 11 pF 11 pF 1 Polarization Voltage 0 V 0 V 2 Physical Housing Material Stainless Alloy Stainless Alloy Venting Rear Rear Mounting Thread Preamplifier 0 4606 60...

Страница 18: ...ARIZED MICROPHONES DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JDM 9 14 05 ECB 9 14 05 CCL 9 14 05 1 1 2 2 A A B B 60 15 3 64 16 4 36 9 2 50 12 7 4606 60 UNS 11 7 mm 60 UNS THREADS 52 13 2 47 12 0 4...

Страница 19: ...plifier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Страница 20: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 21: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 22: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Страница 23: ...ponse reactions of the human ear to the pressure fluctuations Some examples of typical sound pressure levels are shown in the table below dB Pressure Example 0 0 00002 Pa Threshold of Hearing 60 0 02...

Страница 24: ...and noise cones the fre quency response of the microphone will be attenuated at higher frequencies Clamps and stands When holding a microphone in place it is important to minimize the influence of the...

Страница 25: ...2LA manufacturers Our internal metrology laboratory is certified to ISO 9001 accredited by A2LA to AN SI IEC 17025 and ANSI NCSL Z540 1 complies with ISO 10012 1 and former MIL STD 45662A and uses equ...

Страница 26: ......

Страница 27: ......

Отзывы: