For  these  reasons,  it  is  recommended  that  a  recalibration 
cycle be established for each accelerometer.  This schedule 
is unique and is based on a variety of factors, such as: extent 
of  use,  environmental  conditions,  accuracy  requirements, 
trend  information  obtained  from  previous  calibration 
records,  contractual  regulations,  frequency  of  “cross-
checking”  against  other  equipment,  manufacturer 
recommendation,  and  any  risk  associated  with  incorrect 
readings.    International  standards,  such  as  ISO  10012-1, 
provide  insight  and  suggest  methods  for  determining 
recalibration  intervals  for  most  measuring  equipment.  
With  the  above  information  in  mind and under “normal” 
circumstances,  PCB  conservatively  suggests  a  12-  to  24-
month  recalibration  cycle  for  most  piezoresistive 
accelerometers.   
 

NOTE:

 

It  is  good  measurement  practice  to  verify  the 

performance  of  each  accelerometer  with  a  calibration 
device before and after each measurement. 

 

 
 
8.1 RECALIBRATION SERVICE 
 

PCB  offers  recalibration  services  for  our  piezoresistive 
accelerometers,  as  well  as  units  produced  by  other 
manufacturers.    Our  internal  metrology  laboratory  is 
certified  to  ISO  9001,  accredited  by  A2LA  to  ANSI/IEC 
17025  and  ANSI/NCSL  Z540-1,  complies  with  ISO 
10012-1  (and  former  MIL-STD-45662A),  and  uses 
equipment directly traceable to NIST.  Our investment in 
equipment,  traceability  and  conformance  to  industry 
standards  ensures  accurate  calibration  against  relevant 
specifications, in a timely fashion. 

 
 
 
8.2 BACK-TO-BACK CALIBRATION THEORY 

 
Many  companies  choose  to  purchase  the  equipment 
necessary to perform the recalibration procedure in house.  
While  this  may  result  in  both  a  savings  of  time  and 
money,  it  has  also  been  attributed  to  incorrect  readings 
and  costly  errors.    Therefore,  in  an  effort  to  prevent  the 
common  mistakes  associated  with  customer-performed 
calibration,  this  document  includes  a  broad  overview  of 
the  Back-to-Back  Calibration  technique.    This  technique 
provides  a  quick  and  easy  method  for  determining  the 
sensitivity  of  a  test  accelerometer  over  a  wide  frequency 
range.   
 
Back-to-Back  Calibration  is  perhaps  the  most  common 
method  for  determining  the  sensitivity  of  piezoresistive 
accelerometers.    This  method  relies  on  a  simple 
comparison  to  a  previously  calibrated  accelerometer, 
typically referred to as a reference standard.   
 

 

Figure 9.

 Reference Standard Accelerometer 

 
These high-accuracy devices, which are directly traceable to 
a recognized standards laboratory, are designed for stability, 
as  well  as  configured  to  accept  a  test  accelerometer.    By 
mounting a test accelerometer to the reference standard and 
then  connecting  this  combination  to  a  suitable  vibration 
source,  it  is  possible  to  vibrate  both  devices  and  compare 
the  data  as  shown  in  Figure  10.    (Test  set-ups  may  be 
automated and vary, depending on the type and number of 
accelerometers being calibrated). 
 

Figure 10.

 Typical Back-to-Back Calibration System 

 
Because the acceleration is the same on both sensors, the 
ratio  of  their  outputs  (V

T

/V

R

)  must  also  be  the  ratio  of 

their  sensitivities.    With  the  sensitivity  of  the  reference 
standard  (S

R

)  known,  the  exact  sensitivity  of  the  test 

sensor  (S

T

)  is  easily  calculated  by  using  the  following 

equation: 
 

 

S

= S

R

 (V

T

/V

R

 
By varying the frequency of the vibration, the sensor may 
be  calibrated  over  its  entire  operating  frequency  range.  
The  typical  response  of  an  unfiltered  accelerometer  is 
shown in Figure 11. 

 

Figure 11.

 Typical Test Accelerometer Response 

 
8.3 PCB CALIBRATION PROCEDURE 

Numerous precautions are taken at PCB to insure accurate 
and  repeatable  results.    This  section  provides  a  brief 
overview of the primary areas of concern. 

Содержание 3741B122G

Страница 1: ...Accelerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Страница 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Страница 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Страница 6: ...he applied acceleration The sensor element is electrically connected in a Wheatstone bridge circuit The integral amplifier provides a low impedance output to follow on instrumentation These particular accelerometers provide a dc or steady state acceleration measurement capability as opposed to a predetermined low frequency cut off associated with charge or ICP type accelerometers Examples where Hi...

Страница 7: ...assembly With each conductor being 30 AWG Teflon coated The supplied cable length for a standard DC accelerometer is 10 ft or 120 in 3 m Custom cable length requirements should be specified at time of ordering When applications involve very long cable runs the user is advised to take into consideration the added resistance of the cable wires and additional capacitance values as this may affect the...

Страница 8: ... and stiff adhesives provide the best high frequency response METHOD 1 Adhesive Mounting Base This method involves attaching a base to the test structure then securing the sensor to the base This allows for easy removal of the accelerometer Contact the PCB factory representative for optional mounting plate details STEP 1 Prepare a smooth flat mounting surface A minimum surface finish of 32 in 0 00...

Страница 9: ...les to reduce the effects of EMI and RFI that is present near electrical motors and machinery Teflon jacketed cabling may be necessary to withstand corrosive environments and higher temperatures Feel free to contact a factory representative for a specific recommendation on cables that are best suited for your application STEP 2 Connect the cable to the accelerometer A small amount of thread lockin...

Страница 10: ...nits provide the capability for continuous monitoring Note High sensitivity DC Accelerometers feature on board differential amplification 7 0 OPERATING NOTE Always operate the accelerometer within the limitations listed on the enclosed Specification Sheet Operating the device outside these parameters can cause temporary or permanent damage to the sensor 8 0 ACCELEROMETER CALIBRATION Accelerometer ...

Страница 11: ...gs and costly errors Therefore in an effort to prevent the common mistakes associated with customer performed calibration this document includes a broad overview of the Back to Back Calibration technique This technique provides a quick and easy method for determining the sensitivity of a test accelerometer over a wide frequency range Back to Back Calibration is perhaps the most common method for d...

Страница 12: ...ously calibrated over the frequency and or amplitude range of interest 2 the shaker should be selected to provide minimal transverse lateral motion and minimal distortion and 3 the quality of the meters signal generator and other devices should be selected so as to operate within the limits of permissible error 8 4 COMMON MISTAKES Most calibration errors are caused by simply overlooking some of th...

Страница 13: ...return policies and instructions When unexpected measurement problems arise call our 24 hour SensorLineSM to discuss your immediate dynamic instrumentation needs with a factory representative Dial 716 684 0001 3425 Walden Avenue Depew NY 14043 2495 USA Vibration Division toll free 888 684 0013 24 hour SensorLineSM 716 684 0001 FAX 716 685 3886 E mail sales pcb com Website www pcb com ISO 9001 CERT...

Страница 14: ......

Страница 15: ......

Страница 16: ......

Отзывы: