background image

4

pins.) In many cases, it is also helpful to protect the
solder joint with heat-shrink tubing or epoxy.

Note:

If you do not have the experience or resources to

attach pigtail leads, consult PCB to discuss factory
attachment. Damage caused by excessive heat is not
covered by warranty.

STEP 3:

Route the cable to the signal conditioner,

making

certain

to

stress-relieve

the

sensor/cable

connection and minimize motion by clamping the cable
at regular intervals.

Certain precautions should be used to avoid physical
damage and minimize electrical noise.

For instance,

avoid routing cables near high-voltage wires. Do not
route cables along floors or walkways where they may
be stepped on or become contaminated. Shielded cables
should have the shield grounded at one end only.

STEP 4:

Finally, connect the remaining cable end to the

signal conditioner. To dissipate charge that may have
accumulated in the cable, short the signal lead to the
ground lead prior to attachment.

6.0 POWERING

A conventional method for conditioning the high-
impedance signal generated by a charge output sensor is
to use a laboratory-style charge amplifier. The charge
amplifier converts the high-impedance charge signal
generated by the sensor to a low-impedance voltage
signal that can then be transmitted to a readout or a
recording device for analysis.

Before connecting the

low-noise cable from the accelerometer to the charge
amplifier, be sure to ground both the charge amplifier
and the cable.

This ensures that an excessive static

charge

that

may

have

accumulated

across

the

accelerometer or cable is harmlessly discharged. Failure
to observe this precaution can result in the destruction of
the input FET of certain amplifiers.

Once each element is connected, allow a few minutes for
the system to thermally stabilize. Place the switch on
the charge amplifier in the OPERATE position and
proceed with the measurement.

It is often convenient to normalize the accelerometer and
charge amplifier system to a precise sensitivity, such as
10.0 or 100.0 mV/g for ease of data analysis. This is
accomplished

with

most

PCB

laboratory

charge

amplifiers and some miniature in-line units as well.

For fixed sensitivity in-line charge converters, such as
the PCB Series 422, the system sensitivity (mV/g) is
determined as the product of the charge amplifier
sensitivity (mV/pC) and the charge sensitivity of the
accelerometer (pC/g).

Note:

When using charge-amplified systems, the noise

floor of the system is dependent on the input
capacitance to the charge amplifier. Since the cable
adds to the capacitance and to minimize the noise
threshold,

keep

the

cable

length

between

the

accelerometer and the charge amplifier to a minimum.
Cable length does not affect the system sensitivity of
charge-amplified systems.

Since charge amplifier operation varies, please contact
the respective signal conditioner manufacturer or check
the product manual for additional information.

7.0 ACCELEROMETER CALIBRATION

Accelerometer calibration provides, with a definable
degree of accuracy, the necessary link between the
physical quantity being measured and the electrical
signal generated by the sensor. In addition, other useful
information concerning operational limits, physical
parameters, electrical characteristics, or environmental
influences may also be determined. Without this link,
analyzing data becomes a nearly impossible task. PCB
provides a calibration record that documents the exact
characteristics of each sensor. (The type and amount of
data varies depending on the sensor type, contractual
regulations, and other special requirements.)

Under

normal

operating

conditions,

piezoelectric

sensors are extremely stable, and their calibrated
performance characteristics do not change over time.
However,

harsh

environments

or

other

unusual

conditions that cause the sensor to experience dynamic
phenomena outside of its specified operating range may
temporarily or permanently affect the sensor.

This

change manifests itself in a variety of ways, including a
shift of the sensor resonance due to a cracked crystal, or
a temporary loss of low-frequency measuring capability
due to a drop in insulation resistance.

Содержание 357B61

Страница 1: ...lerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mai...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...for most models These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high temperatures preclude the use of accelerometers...

Страница 5: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Страница 6: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Страница 7: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Страница 8: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Страница 9: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Страница 10: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Страница 11: ...Size Hex x Height 5 8 in x 1 0 in 5 8 in x 25 4 mm Weight 1 1 oz 30 gm 1 Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side All specifications are at...

Страница 12: ......

Страница 13: ...ONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M...

Отзывы: