background image

Model Number

357B34

CHARGE OUTPUT ACCELEROMETER

Revision: K

ECN #: 46956

Performance

ENGLISH

SI

Sensitivity(± 15 %)

100 pC/g

10.2 pC/(m/s²)

Measurement Range

± 150 g pk

± 1470 m/s² pk

Frequency Range(+5 %)

3 kHz

3 kHz

[2]

Frequency Range

(+10 %)

3.5 kHz

3.5 kHz

[2]

Frequency Range

(+3 dB)

5.5 kHz

5.5 kHz

[2]

Resonant Frequency

 13 kHz

 13 kHz

Non-Linearity

 1 %

 1 %

[3]

Transverse Sensitivity

 5 %

 5 %

Environmental

Overload Limit(Shock)

± 2000 g pk

± 19,600 m/s² pk

Temperature Range

-95 to +500 °F

-71 to +260 °C

Temperature Response

See Graph

See Graph

[1]

Base Strain Sensitivity

0.0002 g/µ

ε

0.002 (m/s²)/µ

ε

[1]

Electrical

Capacitance

930 pF

930 pF

[1]

Insulation Resistance(at 500°F)

 10

8

 Ohm

 10

8

 Ohm

[1]

Insulation Resistance

(at 70° F [21°C])

 10

12

 Ohm

 10

12

 Ohm

Output Polarity

Negative

Negative

Physical

Sensing Element

Ceramic

Ceramic

Sensing Geometry

Shear

Shear

Housing Material

Titanium

Titanium

Sealing

Hermetic

Hermetic

Size (Hex x Height)

3/4 in x 1.30 in

3/4 in x 33.0 mm

Weight

1.60 oz

45.4 gm

[1]

Electrical Connector

10-32 Coaxial Jack

10-32 Coaxial Jack

Electrical Connection Position

Top

Top

Mounting Thread

10-32 Female

10-32 Female

[4]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

J

- Ground Isolated

Frequency Range(+5 %)

3 kHz

3 kHz

Frequency Range

(+10 %)

3.5 kHz

3.5 kHz

Resonant Frequency

 12 kHz

 12 kHz

Electrical Isolation(Base)

>10

8

 Ohm

>10

8

 Ohm

Size - Hex x Height

0.75 in x 1.35 in

19.1 mm x 34.3 mm

P

- Positive Output Polarity

Output Polarity

Positive

Positive

W

- Water Resistant Cable

Electrical Connector

Sealed Integral Cable

Sealed Integral Cable

Electrical Connection Position

Top

Top

NOTES:

[1] Typical.
[2] Low frequency response is determined by external signal conditioning electronics.
[3] Zero-based, least-squares, straight line method.
[4] See PCB Declaration of Conformance PS160 for details.

SUPPLIED ACCESSORIES: 

Model 080A109 Petro Wax (1)
Model 081B05 Mounting Stud (10-32 to 10-32) (1)
Model ACS-1 NIST traceable frequency response (10 Hz to upper 5% point). (1)
Model M081B05 Mounting Stud 10-32 to M6 X 0.75 (1)

Entered: LK

Engineer: BAM

Sales: WDC

Approved: BAM

Spec Number:

Date: 8/8/2017

Date: 8/8/2017

Date: 8/8/2017

Date: 8/8/2017

357-2340-80 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

Содержание 357B34

Страница 1: ...accel 100 pC g 3k Hz 10 32 top Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 F...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Страница 7: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Страница 8: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Страница 9: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Страница 10: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Страница 11: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Страница 12: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Страница 13: ...erest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical spe...

Страница 14: ...ONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M...

Страница 15: ......

Отзывы: