background image

 

 

 

Manual Number: 43331 

 

                                                                                                     Manual Revision: B 

 

 

                                                                                                       

 

 

3500/3600 Series Piezoresistive Shock Accelerometer Operating Guide 

 

 3 

the shunt resistor are accurately known. Among other sources 
of  variability,  temperature  can  change  the  value  of  the 
resistors, and therefore the Shunt Calibration output. The user 
should  be  aware  that  because  the  sensors  are  full  bridge 
transducers,  all  legs  are  active  and  will  have  temperature 
coefficients  typical  of  piezoresistive  sensors  (approximately 
+0.1%/degree C). 
 

9.0   

Sensor Calibration 

 
For  shock  accelerometers,  the  most  appropriate,  accurate  and 
reliable  calibration  is  with  a  comparison  shock  using  a  back-
to-back  reference  and  pneumatic  exciter,  as  described  in  ISO 
16063-22, “Methods for the calibration of vibration and shock 
transducers  —  Part  22:  Shock  calibration  by  comparison  to  a 
reference transducer.” PCB offers this calibration as a service.

 

 

10.0  

Maintenance and Repair 

 

Because  of  the  sophisticated  nature  of  PCB  instrumentation, 
field  repair  of the equipment is not recommended.  Most PCB 
sensors are of modular construction and are factory repairable. 
A  repair  or  replacement  quotation  is  available  at  no  charge. 
Before returning equipment for repair, it is recommended that 
the  user  confer  with  a  factory  application  engineer  (or  inter-
national representative) to first troubleshoot the problem. 
 

11.0 

Return Procedure 

 

To  expedite  the  repair  process,  contact  a  factory  application 
engineer  to  obtain  a  Return  Material  Authorization  (RMA) 
number prior to sending equipment to the factory. Please have 
information,  such  as  model  number,  serial  number  and 
description of the problem, available. 
 
Customers  outside  the  U.S.  should  consult  their  local  PCB 
distributor  for  information  on  returning  equipment.  For 
exceptions  to  this  guideline,  please  contact  the  International 
Sales department to request shipping instructions and an RMA. 
 
For further assistance, please call (716) 684-0001 or fax us at 
(716) 684-0987. You may also receive assistance via e-mail at 

[email protected]

 or visit our web site at

 

www.pcb.com

 

 

12.0  Customer Service / Warranty 
 

The  employees  of  PCB  strive  to  provide superior, unmatched 
customer  service.  Should  you  at  any  time  find  yourself 
dissatisfied  with  any  PCB  product  for  any  reason,  consult  a 
factory Application Engineer or local representative/distributor 
to discuss repair, refund, or exchange procedures. 
 
When  unexpected  measurement  problems  arise,  call  our  24-
hour  Sensor  Line

TM

  at  (716)  684-0001  to  discuss  your 

immediate  dynamic  instrumentation  needs  with  a  Factory 
Representative. 

Содержание 3503C2020KG

Страница 1: ...riaxial Shock Accelerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Страница 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...icro electromechanical systems MEMS technology in which all structural and electronic components are manufactured in silicon using electronic microfabrication technologies MEMS devices are preferred i...

Страница 7: ...ck the mounting torque after each measurement It s also recommended that a thread locking adhesive be applied to any threads if applicable Ceramic LCC leadless chip carrier packages available in this...

Страница 8: ...ipment for repair it is recommended that the user confer with a factory application engineer or inter national representative to first troubleshoot the problem 11 0 Return Procedure To expedite the re...

Страница 9: ...g Hermetic Hermetic Size Height x Length x Width 0 15 in x 0 38 in x 0 28 in 3 8 mm x 9 6 mm x 7 1 mm Weight 0 029 oz 0 82 gm 4 Electrical Connector Solder Tabs Solder Tabs Mounting Surface Mount Surf...

Страница 10: ...2 DEGREES FILLETS AND RADII 003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 58195 1 OF 1 5X MODEL 3503C20xx SERIES SMT TRIAXIAL ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 7 2...

Отзывы: