9.0    TROUBLESHOOTING 

 

When    a    PCB    signal   conditioner  with   any   of  the 

following    indicators  are   used,   tum  the power on and 

observe the voltmeter ( or LED's) on the front panel. 

  

 

 

Output    voltage   moves   from   YELLOW   to   GREEN 

slowly   until    charging   is  complete. AC coupled  signal 

conditioners   require    sufficient    time     to    charge  the 

internal    coupling  capacitor. Allow signal conditioner  to 

charge   for  approximately 5 discharge time constants  for 

stable operation

 

Note : 

Most  PCB  force  sensors  have an output bias of 8- 

14   VDC.    Refer   to   the   specification  sheet  with  this 

manual  for  the bias  range  of the model you are using. If 

you   are  using  a  low  bias sensor, the indicator will be at 

the  bottom end   of the  green  portion of the dial indicator, 

and may even be in the red portion. This is the expected 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10.0   MAINTENANCE 

 

The  sensor connector must be kept clean, especially if it 

is  operating in a dusty and

/

or wet environment. Because 

the  force  sensor  is  of welded construction

it should be 

returned   to   the   factory   for   servicing in the event of 

serious malfunction

 

Observe the following precautions in using the sensor: 

 

■ 

  

Do   not   exceed   the   maximum  load 

levels 

for the 

     force sensor (see specification sheet)

 

■ 

  

Do  not subject the sensor to temperatures exceeding 

     that of the specification, normally 250°F (121°C). 

 

■ 

  

Do   not apply voltage to the sensor without current- 

     

limiting 

diodes or other current protection.

 

 

■ 

  

Do   not   apply   more  than 20 mA of current to the 

     force sensor. 

 

■ 

 

When     mounting      the      force   sensor,   observe  

     installation procedures detailed in Section 3

.

0 and as     

     outlined  on  the  specific  sensor  installation/outline     

     drawing to avoid overtorquing when mounting. 

 

  

Do not apply more than 30 volts to the sensor.

 

 

  

Avoid   metal-to-metal   impacts   during applications,

 

    which     can       produce     high   frequency    ringing.  

    Electrical  low-pass   filtering   or  a damping material 

    can help reduce such effects. 

 

   

Do    not   spin   the   sensor   onto the cable. This may 

     fatigue     the    cable   pins,   causing   cable   damage. 

     Always   insert   the    cable    pin   into the sensor and 

     tighten  the knurled cable nut to the sensor. 

 

 

 

 

 

 

 

 

 

 

   ICP® is a registered trademark of PCB Piezotronics

Inc. 

 

Содержание 261B03

Страница 1: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 2: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 3: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 4: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 5: ...3 6 0 Low Frequency Monitoring 3 7 0 Discharge Time Constant 4 8 0 Calibration 4 9 0 Troubleshooting 5 10 0 Maintenance 5 Figure Page 1 3 Component Force Link Axis Definition 2 2 ICP 3 Component Force...

Страница 6: ...sed upon the sensor ICP force link sensors contain built in microelectronic signal conditioning circuitry to provide clean low impedance output signals that can be transmitted over low cost cables and...

Страница 7: ...m 2 to 20 mA PCB standard battery powered signal conditioners are factory set at 2 mA and may be used to adequately drive a signal for 100 feet PCB line signal supplies are factory set at 4 mA and adj...

Страница 8: ...DTC represents the amount of time taken for the signal to decay to 37 of the initial peak value As illustrated in Figure 4 this is an exponential decay Approximately five DTC intervals are needed for...

Страница 9: ...sensor is of welded construction it should be returned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor Do not exceed the maximum...

Страница 10: ...ectral Noise 1 Hz Z 0 02 lb Hz 0 089 N Hz Spectral Noise 10 Hz Z 0 005 lb Hz 0 0222 N Hz Spectral Noise 100 Hz Z 0 002 lb Hz 0 0089 N Hz Spectral Noise 1 000 Hz Z 0 0005 lb Hz 0 00222 N Hz Output Pola...

Страница 11: ...ENGTH 5 2 0 5 LENGTH 100 6 0 100 LENGTH 1 0 2 54cm LENGTH 30 5cm 2 54cm 0 30 5cm LENGTH 1 5m 5 1cm 0 1 5m LENGTH 30 5m 15 2cm 0 30 5m LENGTH 30 5cm 0 A A Y X Z 74529 PCB Piezotronics Inc claims propri...

Отзывы: