background image

 

 

 

 

 

ICP

 FORCE SENSOR OPERATION MANUAL 

 

 3 

 

Figure 5 - Series 208 General Purpose and Axial 

Connector ICP

® 

Force Sensor Installation Methods 

 

LINKS 

Series 221B to 227C ICP

  Link Sensors are designed for 

measuring compression from a few pounds(N) to 50,000 lbs (to 

222.4 kN), and tension forces from a few pounds (N) to 30,000 

lbs (to 133.4 kN).  A link consists of a standard PCB ring sensor 

preloaded between two hex end nuts.  All hex nuts are internally 

threaded for mounting ease to a customer test structure.  Unlike 

ring design sensors, additional  external preloading is  not 

required  as  the factory installed  external mounting hardware 

places the sensor in a preloaded state. 

 

The hex nuts do not loosen naturally.  

DO NOT ADJUST THE 

NUTS WITH A WRENCH. 

 Loosening or tightening of the 

hex nuts will change the preload applied to  the sensor. The 

result is that the sensor output will no longer match the factory 

supplied calibration certificate.  In the unlikely event the 

mounting hardware becomes loose, contact PCB’s service and 

repair document for proper instructions for returning the sensor 

to PCB for recalibration.. 

 

Figure 6 outlines some possible mounting configurations of the 

link series of sensors. 

 

 

 

Figure 6 – Series 221 to 227 ICP

®

 Force Link Sensor 

Mounting Method 

 

 

3- COMPONENT TRIAX SENSORS

 

PCB’s line of 3-Component force sensors is  capable of 

simultaneously measuring force in three  (3)  orthogonal 

directions (X, Y, and Z).  Internally, three sets of quartz crystals 

are cut, oriented and preloaded so the maximum  output from 

the crystal cut for the specific plane is obtained when a force is 

applied axially to the specific x, y or z plane.  Sensor output 

from loads applied in the alternate, transverse 

planes(considered channel cross-talk) is  negligible as the 

crystal sets are cut for maximum output in a specific plane. 

 

Optimum performance and linear operation of 3-component 

force sensors is obtained when a preload is applied to the sensor.  

Versions are available with ranges up to 10K lb (45kN) in the 

Z-axis (perpendicular to the top surface), and up to 4,000 lbs 

(18kN) in the X and Y (shear) axes. 

 

There are two modes of triaxial force sensors.   

ICP

 triaxial 

designs utilize built-in microelectronic circuitry that provides a 

low-impedance voltage, the electrical output of which is passed 

to external signal conditioning  via  one  multi-pin connector 

mounted on the sensor housing  This design mechanically has 

one 4-pin electrical connector that may be coupled to a  single 

multi-conductor sensor cable.  

  

High  impedance

  charge  mode models contain three  (3) 

separate electrical output  connectors  on the sensor housing, 

each corresponding to the respective x, y, or z axis.  Low noise 

cables should be used in low impedance system arrangements.  

These models are ideal for use in harsh industrial  or high 

temperature environments. 

 

Figure 7 outlines installation possibilities of triax force sensors.   

 

Содержание 209C11

Страница 1: ... 2 2 lb comp 1 lb tension 2200 Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...stallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description of the various sensor series available recommended mounting procedures operation and recommended calibration In addition to PCB s line of ICP sensors each ICP sensor series outlined has corresponding ...

Страница 7: ...e ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Installation Methods GENERAL PURPOSE RADIAL Model 208C01 C05 General Purpose Sensors are designed to measure compression and impact forces from a fraction of a lb N to 5 000 lbs to 22 24 kN Tension force...

Страница 8: ... simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the specific plane is obtained when a force is applied axially to the specific x y or z plane Sensor output from loads applied in the alternate transverse planes considered channel cross talk is negligible ...

Страница 9: ...returning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for compression and impact force measurements in materials testing applications Smooth cylindrical housings and curved impact caps avoid cutting through specimens This enables measurements to be taken to de...

Страница 10: ...ot facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surfaces are flat and parallel Non parallel surfaces could place unwanted stress on internal components leading to premature sensor failure The protective cap should remain on the electrical connector durin...

Страница 11: ...the left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total force for the beryllium copper stud supplied with the sensor and up to 50 for steel studs This typical installation setup is used by PCB during standard calibrations of force ring models 201B01 through ...

Страница 12: ...nitor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accomplished in steps not to exceed the usable voltage To prevent clipping of the signal increments should not exceed 10 VDC Figure 14 ICP Force Sensor Preload Requirements WORKING RANGE PRELOAD AND MAXIM...

Страница 13: ... hundreds of feet meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent circuit contained in all PCB signal conditioners See Guide G 0001B for powering and signal conditioning information pertaining to all ICP instrumentation Most of the line powered signal conditioners ...

Страница 14: ...tant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents the decay rate of an input signal One DTC represents the amount of time taken for the signal to decay to 37 of the initial peak value As illustrated in Figure 16 this is an exponential decay Approxim...

Страница 15: ...turned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specification sheet B Do not subject the sensor to temperatures exceeding that of the specification normally 250 F 121 C C Do not apply voltage to the sensor without current limiting diodes or other current pr...

Страница 16: ...el Stainless Steel Sealing Hermetic Hermetic Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side Mounting Thread 10 32 Female No Metric Equivalent All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered tradem...

Страница 17: ......

Отзывы: