PCB Piezotronics 123 Series Скачать руководство пользователя страница 7

 

Drawing Number: 21091 

 

ECO: 52570 

Revision: B 

 

OPERATION MANUAL FOR 

HELIUM BLEED WATER-COOLED  

PRESSURE SENSORS 

123 Series

 

 3 

level to zero out quickly without effect on the 

dynamic level indicated. 

 

Also, the influx of large amounts of thermal energy, 

as in rocket combustion measurements, can cause the 

sensor to generate spurious transient outputs of 

considerable magnitude.  Using the charge amplifier 

on "SHORT" TC alleviates this problem by shunting 

these longer term charge outputs to ground while 

measuring the faster dynamic pressure pulses. 

 

4.3 

OPERATION - ICP

 MODELS  

 

Consult Guide G-0001B for a complete description of 

the low-impedance concept in instrumentation. 

All PCB power units designed for ICP

 use contain 

built-in fault monitor meters to aid in locating circuit 

faults (shorts or opens) and to identify normal 

operation of the miniature impedance-converting 

amplifier built into the sensor. 

 

Do not attempt to apply voltage directly to the input 

pin of the sensor from any other power source.  The 

input current must be limited to 20 mA maximum and 

the input voltage must not be higher than 28 VDC. 

Consult the factory if it is desired to build a power 

unit rather than to use one of those available from 

PCB. 

 

NOTE:  Full-scale range is 1imited by the helium 

source pressure.  The applied helium source pressure 

level must be at least twice the peak combustion 

pressure (not to exceed 15,000 psi), and not more 

than 10 times the lowest expected peak combustion 
pressure. This is true for all models, charge or ICP

 

5.0 

CALIBRATION 

 

5.1 

CHARGE MODEL

 

 

To calibrate this model, install in pressure port of 

dead weight tester and apply various static pressures, 

recording corresponding outputs at each pressure of 

interest. 

 

Discharge the charge amplifier before each reading 

and return pressure to zero before each new pressure 

input.  Use the charge amplifier on “LONG” time 

constant and keep cables and connections clean to 

avoid drift.  A DC digital voltmeter is ideal for 

readout during calibration. 

NOTE:   The helium tube must be plugged with a 

pressure-tight seal before attempting to calibrate to 

avoid losing pressure through the helium tube. 

 

If a hydraulic calibration system is used, be sure to 

remove all traces of oil from the diaphragm passage 

after calibration.  Pneumatic systems are preferred for 

calibration. 

 

5.2 

ICP

 MODELS

 

 

Because of the shorter time constants built into these 

models, static calibration methods are much more 

difficult to use.  Pressures must be applied rapidly 

and readout must be made with fast recording devices 

such as light beam oscillographs or storage 

oscilloscopes. 

 

A pneumatic (compressed air or hydrogen) system 

with quick opening valve and reference gage is an 

excellent method of calibrating these sensors. 

 

MAINTENANCE

  

 

The sealed construction and miniature size of these 

sensors make field repair impractical. 

 

The helium filter element should be cleaned should 

the flow rate  decrease.  Remove the element and 

ultrasonically agitate in solvent for 15 minutes, then 

dry. 

 

Should the orifice become plugged, back pressure the 

orifice by applying air pressure to diaphragm end of 

sensor with filter off and helium tube vented to 

atmosphere. 

 

The only maintenance possible on the probe is 

maintaining of insulation resistance of the charge 

mode models.  Clean connectors with suitable solvent 

and dry. 

 

Contact factory for assistance should problems arise. 

 

 

 

 

 

 

ICP is a registered trademark of PCB Piezotronics 

Содержание 123 Series

Страница 1: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 2: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 3: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 4: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 5: ...ut tube The purpose of this filter is to prevent particles in the helium supply from plugging the metering orifice The outer body of this series is 17 4 PH hardened stainless steel and features a ceramic coated tip for protection against erosion by hot gasses when the sensor is flush mounted directly in combustion chambers The probes in all models are electrically ground isolated from the outer ho...

Страница 6: ...t panel monitor meter to check cable connections and internal amplifier Consult Guide G 0001B General Guide to ICP Instrumentation for installation and usage hints for ICP instruments 4 0 OPERATION 4 1 CHARGE OUTPUT MODEL For the charge mode it is only necessary to select the proper range on the charge amplifier and proceed with the measurement The probe sensitivity is approximately 1 0 pC psi The...

Страница 7: ... tester and apply various static pressures recording corresponding outputs at each pressure of interest Discharge the charge amplifier before each reading and return pressure to zero before each new pressure input Use the charge amplifier on LONG time constant and keep cables and connections clean to avoid drift A DC digital voltmeter is ideal for readout during calibration NOTE The helium tube mu...

Страница 8: ...inless Steel Diaphragm Invar Invar Sealing Epoxy Epoxy Electrical Connector Integral Cable Integral Cable Weight 4 59 oz 130 gm Helium Flow Rate at 1 500 psi 40 scf hr 1 13 m hr Cable Length 4 ft 1 2 m Water Flow Rate at 50 psi 1 2 gal min 4 5 L min 5 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right tochange ...

Страница 9: ...NTING HOLE PREPARATION M14x1 25 6H THRU 625 001 15 88 02 060 1 52 MOUNTING HOLE PREPARATION 100 2 54 THRU M14x1 25 6H 720 18 29 625 001 15 88 02 060 1 52 1 1 1 2 3 4 SEAL SURFACE MUST BE FLAT AND FREE OF TOOL MARKS WITH A 63 1 5 FINISH RECOMMENDED MOUNTING TORQUE ON 1 125 28 58 HEX 125 5 inch lb 14 5 Nm SUPPORT 375 9 53 HEX ON FILTER WHEN ATTACHING LINE TO AVOID DAMAGE TO HELIUM INPUT TUBE RECOMME...

Отзывы: