Drawing Number: 21086 
Revision: NR

 

 

 

 

OPERATION MANUAL FOR 

CONFORMAL BALLISTICS PRESSURE SENSOR 

Series 117B

 

1.0 

INTRODUCTION 

 

The  Model  117B  Conformal  Sensor  measures 
pressure  inside  metallic  cartridges  and  paper  or 
plastic  shotshell,  during  firing  of  the  round,  without 
drilled holes in the round. 
 
The  sensor is mounted flush with the chamber walls, 
matching  the  curvature  exactly.  The  expanded 
cartridge or shotshell transmits forces through the thin 
wall to the sensor. 
 
The  Model  117B  is  a  conventional  charge  output  
type sensor using synthetic quartz as the piezoelectric 
sensing material.  
 
An  alignment  guide  now  standard  on  all  117B 
Models,  simplifies  installation  by  automatically 
controlling rotational alignment of the sensor, leaving 
only the depth adjustment consideration. 
 

2.0 

DESCRIPTION 

 

Refer  to  enclosed  installation  drawing  for  outline 
dimensions and physical description.  
 
The  basic  sensor  is  a  probe  design  with  captivated 
hollow 

clamp 

nut. 

The 

clamp 

nut 

rotates 

independently of the probe body and serves to lift the 
sensor out of the mounting port as it is unscrewed.  
 
The  alignment  device  consists  of  a  pin  and  slotted 
clamp  assembly  which  fits  tightly  over  the  collar  of 
the sensor.  
 
The hardened  steel  pin is accurately aligned with the 
axis of the diaphragm curvature and when assembled 
into  a  precisely  drilled  mating  hole  in  the  barrel  or 
calibration  chamber,  achieves  near  perfect  rotational 
alignment of sensor diaphragm. 
 
 
The slotted clamp arrangement allows for tolerance in 
the location of the guide pin hole in an axial direction 
on the test barrel or calibration adaptor.  
 
The  precise  depth  adjustment  is  obtained  by  the  use  
of the correct thickness spacer selected from a set of 9 
spacers  of  various  thicknesses  supplied  with  each 
sensor. 

Once  the  proper  thickness  spacer  is  found,  removal 
and re-installation now becomes a routine matter. 
 

3.0  

INSTALLATION 

 

3.1  

PORT PREPARATION 

 
Prepare mounting port in accordance with installation 
drawing  117-20  (XX)-90  supplied  as  part  of  this 
manual. 
 
NOTE: The installation port for the 117B is identical 
to that for the 117A. 
 
Drill guide pin hole as shown in installation drawing, 
paying  particular  attention  to  locating  C/L  of  hole 
exactly on C/L of barrel. 
 
Do not use a drill that is worn, as this may cause the 
hole  to  be  slightly  undersized and the guide pin may 
bind as it is drawn into the hole. 
 
It  is  extremely  important  that  the  guide  pin  hole  be 
drilled parallel to the sensor mounting port  to permit 
the guide pin to move freely into the hole. 
 
NOTE: Each model variation of the 117B is designed 
to  fit  in  a  specific  location  (measured  from  the  bolt 
face)  on  a  specific  ammunition  caliber.  Changing 
location  and/or  caliber  will  cause  a  mismatch  of 
diaphragm  curvature  with  cartridge  case  diameter 
since most cartridge cases are tapered. 
 
 

3.2 

INSTALLING THE SENSOR 

 

After  mounting  holes  have  been  prepared,  proceed 
with installation as follows: 
 
1. 

On most models of 117B, it is not important 
which  side  of  the  sensor  is  mounted  toward 
the muzzle. 

 

However,  on  certain  types  of  ammunition 
(such as 20 mm cannon) which have a severe 
taper, the diaphragm has a matching taper in  

Содержание 117B205

Страница 1: ...PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...ngement allows for tolerance in the location of the guide pin hole in an axial direction on the test barrel or calibration adaptor The precise depth adjustment is obtained by the use of the correct thickness spacer selected from a set of 9 spacers of various thicknesses supplied with each sensor Once the proper thickness spacer is found removal and re installation now becomes a routine matter 3 0 ...

Страница 7: ...ect a thinner spacer and remount Once the proper thickness is found for perfect flushness the sensor may be removed and reinstalled using this same spacer and the proper depth will be achieved each time NOTE For best accuracy of results use same charge amplifier for calibration and for actual operation Use long TC for calibration medium or short TC for best drift free operation 4 0 POLARITY Polari...

Страница 8: ...r to allow easy insertion and extraction the case must be expanded slightly by the internal pressure before force can be transmitted to the sensor See the enclosed guide An Improved Technique for Utilization of Conformal Ballistics Sensor Calibration Data for methods of dealing with this topic 6 0 MAINTENANCE It is essential for normal operation of the Model 117B that the insulation resistance be ...

Страница 9: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 4 ...

Страница 10: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 5 ...

Страница 11: ...mmunition under test to take into account transmissibility characteristics of the cartridge case for each lot of ammunition Because a certain amount of pressure is required to fully obturate the cartridge case in the test chamber a rather severe non linearity is exhibited over the first several thousand psi of input pressure i e until the cartridge case is pressed firmly against the inside chamber...

Страница 12: ...or two percent of full scale is specified for quartz sensors An acceptable instrument is then defined as one whose data points all fall within this error band The sensitivity of an instrument so calibrated is simply the full scale output charge or voltage divided by full scale input pressure where F S output is determined by the straight line as defined above input S F output S F y Sensitivit Eq 1...

Страница 13: ...ance between the cartridge case and the chamber wall Obturation of the cartridge occurs in this region However when obturation is complete the sensor output is then linear with pressure from this point up to maximum rated pressure In the past a common procedure has been to calibrate the conformal at the expected nominal pressure deriving a sensitivity at this point e g point Po Qo in figure 4 The ...

Страница 14: ...he data from 20 consecutive calibration runs performed on a conformal sensor yielded a slope of 04 pC psi and an intercept of 7 500 psi determined by constructing the best fitting straight line as described previously Assume also that average expected peak pressure is 25 000 psi Approximate full scale charge output is pC 1000 pC psi 04 psi 25 000 Set range of charge amplifier output to 1000 pC vol...

Страница 15: ... can be further refined to include a variable voltage source as a bias in series with the output of the charge amplifier This voltage bias can be set to simulate the offset pressure and then the true peak pressure can be read directly See figure 6 on the next page Example Assume a 20 round calibration lot and attendant graph yield a slope of 250 pC psi and an offset of 5 000 psi Proceed as follows...

Страница 16: ...scale pressure to find full scale picocoulombs 40 000 psi x 250 pC psi 10 000 pC Set the charge calibrator for 10 000 pC and apply this signal to the input of the amplifier Now adjust the gain of the oscilloscope or charge amplifier if gain adjust is available to make full scale equal to a selected number of oscilloscope divisions For example 6 Div 40 000 psi Now the peak value can be determined f...

Страница 17: ...Drawing Number 21210 Revision NR AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA 7 Adds offset pressure not available from PCB ...

Страница 18: ...xy Epoxy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 45 oz 13 gm All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as ...

Страница 19: ...001 23 HOLD THIS DIMENSION TO OBTAIN FLUSH MOUNT 1 2 3 4 5 2 1 4 REVISIONS REV DESCRIPTION DIN B UPDATED 065A19 NOTE 50698 117 2205 90 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc 5 236 002 5 99 05 806 20 47 250 6 35 46 11 8 89 22 7 1 63 41 4...

Отзывы: