PCB Piezotronics 103B02 Скачать руководство пользователя страница 6

 

Drawing Number: 21082 
Revision: NR 

 

OPERATION MANUAL FOR 

MICROPHONE 

Series 103A 

 3 

 
Refer to Guide G-0001B for tips on power and signal 
utilization. 
 
NOTE:    The  internal  low-noise  amplifier  in  the 
Model  103A,  A02,  A11  and  A12  differ  from  the 
standard  ICP

®

  amplifiers  described  in  Guide  G-

0001B  in  that  the  turn-on  (or  bias)  voltage  is 
nom4 volts. 
 
Proper  turn-on  for  Models  103A02,  A11  and  Al  2 
will not give center scale reading on the fault monitor 
meters  found  on  most  PCB  signal  conditioners  but 
will show a reading of approximately 20% of F.S. on 
these  meters.    This  puts  it  at  the  lower  limit

 

of  the 

green (or normal) area of the meters.  This is normal 
and in no way indicates faulty operation. 

 

4.0 

OUTPUT SIGNAL

 

 
The  output  signal  from  the  103A  Series  is  a  voltage 
proportional  to  input  pressure.    The  polarity  is 
positive-going for increasing input pressure. 
 
The  signal  is  of  high  enough  level  and  low  enough 
output  impedance  to  be  fed  directly  into  most 
oscilloscopes, tape recorders, strip chart recorders, A 
to D converters, etc. 
 
Consult  the  factory  for  information  regarding  galvo 
driving  since  special  amplifiers  are  needed  to  drive 
most high-current galvanometers. 
 
Low-frequency  response  is  limited  by  sensor 
discharge TC (given in the specification sheet) and by 
coupling  time  constants  throughout  the  system.    To 
take full advantage of the TC of the sensor, make sure 
that  coupling  TC's  from  power  unit  to  readout  are  at 
least an order of magnitude longer than the discharge 
TC or use a direct-coupled signal conditioner such as 
the Model 484B.  Consult factory for details. 
 

5.0 

CALIBRATION 

 
The  103A

 

Series  Sensors  must  be

 

dynamically 

calibrated  using  a  known  step  function  or  pulse,  or 
with a microphone calibrator. 
 
 
 
 
 

 
At PCB,  the units are subjected to various calibrated 
steps of pneumatic pressure to obtain a linearity curve 
over the full range of the instruments. 

 

6.0     MAINTENANCE AND REPAIR

 

 
The  small  size  and  sealed  construction  of  the  Series 
103A precludes field maintenance and repair. 
 
Should  the  time  constant  degrade  or  an  abnormality 
appear in the normal bias voltage, bake the unit out in 
a +250° F oven for 1 to 2 hours, then re-test. 
 
If  this  is  not  effective  or  if  other  problems  should 
occur, contact the factory for assistance in tracing the 
problem  or  for  instructions  on  returning  the  unit  for 
repair. 
 

7.0     PRECAUTIONS 

 
l.) 

Do  not  apply  voltage  to  the  unit  without  a 

current-limiting  device  in  the  line  such  as 
incorporated  in  all  PCB  signal  conditioners,  (20  mA

 

maximum) to do so will destroy the internal amplifier. 
 
2.) 

Do  not  subject  these  units  to  temperatures 

higher than 250° F. 
 
3.) 

Do not attempt to disassemble these units as 

to do so voids the warranty. 
 
4.) 

Use  caution  when  removing  units  that  have 

been mounted with epoxies or other strong adhesives.  
Excessive  squeezing  or  torquing  may  damage  the 
units and/or cause a calibration shift. 
 
5.) 

In vibratory environments, secure cables and 

leads to surrounding structures to avoid lead or cable 
damage. 
 
6.) 

Do  not  over  torque  when  using  the  Servo 

clamps  or  the  clamp  nut since  this could damage the 
units. 
 
7.) 

Do  not  overpressure  (see  specs  for 

maximum  pressure),  to  do  so  may  change  the 
calibration. 

 

 
 
®ICP is a registered trademark of PCB Piezotronics 

Содержание 103B02

Страница 1: ...URE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...igh sensitivity Models 103A02 and 103A12 ICP Sound Pressure Sensors 2 0 DESCRIPTION Refer to installation drawing no 103 1010 90 included as part of this manual The 103A Series Sensors utilize a bimor...

Страница 5: ...2 since the radial wire leads of the Model 103A and 103A11 would interfere with the clamp nut The installation drawing gives detailed instructions on preparation of the port for this type of installat...

Страница 6: ...pled signal conditioner such as the Model 484B Consult factory for details 5 0 CALIBRATION The 103A Series Sensors must be dynamically calibrated using a known step function or pulse or with a microph...

Страница 7: ...s Steel 316L Stainless Steel Sealing Welded Hermetic Welded Hermetic Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 11 oz 3 15 gm All specifications are at room temperature unless...

Страница 8: ......

Страница 9: ......

Страница 10: ......

Отзывы: