TRACKING THE ZERO
This function allows you to perform a momentary zero calibration
compensating for the temperature drift of the weight.
When you switch off the transmitter, it automatically returns to the
previous zero calibration.
The maximum weight resettable by this parameter is 2% of the range
of the system.
To disable this feature, use the value 0.
Value Change
0
Control OFF
1
0.5 div/sec
2
1 div/sec
3
2 div/sec
4
3 div/sec
Default: 0
0 band
ZERO BANd
This parameter defines the number of divisions resettable by the
pressure of the front button of zero or by Input 1.
Values: from 0 to 200
Default: 100