Instructions for Operation and Care of Sammons Preston Rolyan Splint Pans

5

Patterson Medical

 • 1000 Remington Blvd • Suite 210 • Bolingbrook, II 60440 

Phone (800) 323-5547 • Canada (800) 665-9200 • Fax (800) 547-4333 • Canada (905)566-9207

 

CHANGING THE OPERATING 

THERMOSTAT

 

On  the  A426112  (SM-1501)  and  the  A426110             
(SM-1502)  the  temperature  is  varied  by  turning  the 
thermostat  knob  clockwise  for  an  increase  in 
temperature and counterclockwise for a decrease in 
temperature. The temperature on low setting should 
approximate 140ºF and on high setting 190ºF.  

Turn  the  splint  pan  off  via  the  controls  on  the  front 
panel and then unplug the unit. Allow the unit and the 
water to cool before draining the water from unit.

Both  models  have  an  operating  thermostat,  which 
allows for temperature modulation and an overheating 
sensor that prevents the pan from heating over 200ºF. 
To replace the operating thermostat on the A426112 
and A426110, pull the knob off the thermostat dail on 
the front of the unit and remove locking nut. Remove 
the four screws holding the front switch plate. Move 
this  front  switch  plate  to  the  side,  allowing  it  to  be 
supported by the wiring harness.

Follow the curled wire to the capillary tube and loosen 
the  nut  that  clamps  the  capillary  tube  to  the  tank. 
Remove  the  capillary  tube.  Pull  the  wires  off  the 
thermostat and remove. To replace, put the wires onto 
the new thermostat and clamp the capillary tube to the 
tank. Screw the switch plate back onto the unit and put 
the locking nut and thermostat knob back in position. If 
there  are  any  problems  at  all,  please  contact 
Sammons Preston. 

REPAIR OF UNIT

Turn the splint pan off via the controls on the front panel 
and then unplug the unit.  Allow the unit and the water to 
cool before draining the water from the unit.

Next, position the unit to allow access to the rear of the 
unit.  Locate  the  fuse  holders  at  rear  of  the  unit  near 
cord. Depress the fuse holder caps with the use of a 
“common” blade or fingers and rotate counterclockwise 
approximately 90º or until cap becomes loose. Remove 
the fuse holder cap, exposing the glass fuse. Remove 
the blown fuse.

Once the old fuse has been removed, replace it with a 
new fuse of the same kind and amperage. Re-install the 
fuse holder cap by depressing and rotating clockwise 
approximately 90º and ensure that the fuse holder caps 
are secured. Refill the splint pan with at least 3 inches of 
water and then plug the unit back into a power outlet. 
Turn the unit on via front control panel. Allow the water 
to warm  up  to the operating  temperature  and  use  as 
required. 

In addition to the thermostat, which can be replaced on 
site, it may be necessary or you may desire to return the 
Sammons Preston Splint Pan for service to Sammons 
Preston. Please call this number and we will be glad to 
help you: (800) 432-2720 

TO CHANGE TEMPERATURE 

SETTINGS

CHANGING THE FUSE