PERFORMING GRIP STRENGTH MEASUREMENTS

1.  [LB/KG]:

 The Jamar Smart Hand Dynamometer is originally set to display in LBs. To change to 

KG, open up the battery compartment and move the frequency switch down to KG. 

2.  Adjust the grip handle to best suit patient comfort and test requirement (See Adjusting Grip Posi-

tion, page 7). 

3.  Press the 

[ON/OFF] 

key to start the instrument. 

4.  Determine which hand will be used. Using the 

[SELECT TEST]

 key, scroll to the “L” (Left) or “R” 

(Right) icon to select the hand to be tested. 

5.  To select the number of test repetitions (up to 5) to be recorded for that hand, press the  

[# OF TRIALS] 

key. Each of the five numbers (1 through 5) will be displayed on the upper line of 

the display. Stop scrolling on the number which represents the number of tests to be taken. 

6.  Have the patient grasp the Jamar Smart Hand Dynamometer gently so that the palm fits comfort-

ably to the rear of the instrument and the fingers curl around the adjustable grip at the front of the 

instrument. 

7.  Press the 

[TEST]

 key to begin the test. Under the present test protocol, start with the first test trial 

and the number “1” will appear and flash at the top of the display, indicating that the first test for 

the chosen hand is ready. Have the patient squeeze the grip and be ready to enter the second trial 

(if test protocol setting has more than 1 trial). 

8.  When the first test is completed, press the 

[TEST]

 key one time to scroll to the next test. Repeat 

the number of trials to complete the test cycle for the chosen hand.

Example: 

The right hand is chosen to perform four tests. Press the 

[SELECT TEST] 

key and scroll to 

the “R” (Right) icon. Press the 

[# OF TRIALS]

 key until the number “4” is displayed. Press the 

[TEST] 

key to begin the test. “R” and “1” will be displayed on the upper line. Have the patient perform the 

first test. Press the 

[TEST]

 key to move the display to “R2”. Have the patient perform the second test. 

Press the 

[TEST]

 key after each test until all four tests have been performed. 

Содержание Jamar Smart

Страница 1: ...Jamar Smart Hand Dynamometer FROM PATTERSON MEDICAL USER S GUIDE Download the Free App...

Страница 2: ...ing the Number of Trials 6 Selecting Pounds or Kilograms 6 Adjusting Grip Position 7 Installing Replacing Batteries 7 Performing Grip Strength Measurements 8 Recording Grip Strength Measurements 9 Cal...

Страница 3: ...f the same test protocol is used each time Up to 5 tests trials can be conducted with each hand without reset AVERAGE STANDARD DEVIATION and COEFFICIENT OF VARIATION are automatically calcu lated Hand...

Страница 4: ...y to start a new test or if a mistake is made during the test Select number of repeats and supports multiple hand rapid exchange testing Automatic calculations of AVERAGE STANDARD DEVIATION and COEFFI...

Страница 5: ...and with three 3 repeats CHOOSING THE HAND TO BE TESTED Press the RESET key to clear any test which might be in progress Then press the SELECT TEST key to step through the available choices Press SELE...

Страница 6: ...hand selection and number of grip repeats to see how these functions work Once test protocol is changed the Jamar Smart Hand Dynamometer retains the current protocol until changed NOTE If the batteri...

Страница 7: ...nt on the lower part of the handle helps prevent the handle from separating from the shaft during normal use INSTALLING AND REPLACING BATTERIES Turn the instrument so the Key Pad Operator faces front...

Страница 8: ...he rear of the instrument and the fingers curl around the adjustable grip at the front of the instrument 7 Press the TEST key to begin the test Under the present test protocol start with the first tes...

Страница 9: ...he test results stored in the Jamar Smart Hand Dynamometer All the test and calculation results can be viewed on the display by repeatedly pressing the TEST key All results will be displayed in the se...

Страница 10: ...20 8 23 74 5 13 9 54 8 10 4 33 8 6 3 L 110 5 16 2 63 5 12 2 50 1 7 3 28 8 5 5 30 34 R 121 8 22 4 78 7 19 2 55 2 10 2 35 7 8 7 L 110 4 21 7 68 17 7 50 1 9 8 30 8 8 0 35 39 R 119 7 24 74 1 10 8 54 3 10...

Страница 11: ...ide reasonable protection against harmful interfer ence in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance w...

Страница 12: ...e present test protocol start with the first test trial and the number 1 will appear and flash at the top of the display indicating that the first test for the chosen hand is ready Have the patient sq...

Отзывы: