When re-fitting the inner detector assembly ensure the external viewing apertures are
positioned above in line with the sensors.
3.3
Purge Air Supply
The unit is equipped with a fitting for connection to a 10mm compressed air flexible hose.
The air supply must be capable of delivering at least 30 ltrs/min at this connection and have
a pressure of 175 to 700 mBar (2.5 to 10 psi).
Fig. 13
To remove lid
unscrew anti-
clockwise
Fixings A and B
External viewing
apertures
Sensors
10mm compressed
air inlet
3mm hex locking screw
Page 12 of 16
D1198 - Issue 1 - 25 September 2013
Series 5000 - Manual 5610