QUICK START GUIDE
SENTIERO DESKTOP OAE AND TYMP SCREENER
(BLUE HOUSING)
100800-02_Sentiero Desktop OAE + Tymp Screener Quick Start Guide_Rev 3
Page 3 of 7
DISTORTION PRODUCT OTOACUSTIC EMISSIONS (DPOAE) SCREENING
Select either the Screening or the Standard button from the start menu and connect the probe to the red connector.
Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making sure that the ear
tip fits tightly into the ear canal.
Select the test ear by pressing either the Right or Left button.
Cartoon mode will start the test showing a cartoon instead of the progress bar.
A calibration will be performed prior to measurement. A green light indicates successful
calibration. After successful ear probe calibration, the test will start automatically.
Probe fit
override for
patients
with PE
tubes or
middle ear
perforations
Progress for
each tested
frequency
Overall test
progress
Noise floor
level