Path medical Sentiero Скачать руководство пользователя страница 2

QUICK START GUIDE 

SENTIERO DESKTOP OAE AND TYMP SCREENER 

(BLUE HOUSING)

 

 

 

100800-02_Sentiero Desktop OAE + Tymp Screener Quick Start Guide_Rev 3 

 

Page 2 of 7 

Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making sure that the ear 
tip fits tightly into the ear canal.  

TYMPANOMETRY TEST SCREEN 

The system will first check the probe fit and wait for the pressure build up. The LED light on the probe will pulse at a 
steady pace to indicate a correct fit. During the test, if the probe fit is not correct, the LED light will blink at a fast 
pace and an error message prompting you to retry or cancel the test will appear on the screen. 

 

 
 
 

                    

 

 

 

 

Once a proper seal has been obtained, the test will proceed. A tympanometry test will be performed, followed by a 
reflex measurement.  

 

 

Tympanometry test 
result 

Overall Reflex test result 

Reflex test in progress 

Reflex threshold found 

Содержание Sentiero

Страница 1: ...he start menu select the Tymp Tymp 1kHz button Select the test ear by pressing either the Right or Left button Cartoon mode will show an airplane on the screen during the tympanometry test For web des...

Страница 2: ...first check the probe fit and wait for the pressure build up The LED light on the probe will pulse at a steady pace to indicate a correct fit During the test if the probe fit is not correct the LED li...

Страница 3: ...on the probe tip Insert the probe in the ear making sure that the ear tip fits tightly into the ear canal Select the test ear by pressing either the Right or Left button Cartoon mode will start the te...

Страница 4: ...EFER You can retest the frequencies that did not pass Print results to a label or to a pdf file or full page report on PC Note These circles indicate that other screens are available to view Swipe you...

Страница 5: ...adphones to the blue connector and select the Audiometry Screening button from the start menu Select the ear on which you want to start the test AUDIOMETRY TEST SCREEN Select the stimulus levels Selec...

Страница 6: ...ing status for a frequency has been identified move on to the next frequency Tapping twice on a square will display NR no response Tap three times to get back to blank square Once all frequencies have...

Страница 7: ...TH medical reserves the right to make changes at any time without notice Please note that not all products displayed on our webpage or advertising materials are available in all territories Please con...