B l o o d   P r e s s u r e   S e n s o r

T r o u b l e s h o o t i n g

6

Troubleshooting

Why are the Systolic and/or Diastolic Pressure Not Appearing?

Perhaps the size of the cuff is incorrect. You may have a cuff that is too small or 
too large. 

The pressure in the cuff may be falling too quickly, or not fast enough. The pres-
sure release rate should be between 2 and 5 mmHg per second, and never less 
than 2 mmHg per second.

Adjusting the Pressure Release Rate for the Blood Pressure Cuff

The default pressure release rate is 3 mmHg per second. To adjust the rate up or 
down, use a small screwdriver to turn the adjustment screw in the center of the top of 
the push-button valve. Turn the screwdriver counter-clockwise to slow down the 
release rate and turn the screwdriver clockwise to speed up the release rate. The pres-
sure release rate is normally slower if the person’s arm is larger.

Storage

Disconnect the blood pressure cuff from the sensor for storage.

Average Blood Pressure Values

Specifications

Measurement

 Range

Normal

120/80 or less

Pre-hypertension

120/80 to 140/90

High

greater than 140/90

Very high

greater than 180/110

Heart Rate

Blood Pressure

Gage Pressure

Units

beats per minute 

(bpm)

millimeters of 

mercury (mmHg)

mmHg, N/m

2

, kPa, 

atm, psi

Range

36 to 200 bpm

0 to 375 mmHg

0 to 375 mmHg

Accuracy

1 bpm

see gage pressure

3 mmHg

Resolution

1 bpm

see gage pressure

0.05 mmHg

Adjust the 

screw on 

the 

push-button 

valve

Mean Arterial Pressure 
(MAP)

Ideal MAP is defined as 
93 mmHg, which corre-
sponds to 120/80.

SP = Systolic Pressure
DP = Diastolic Pressure

Three formulas are used to 
compute MAP. All three pro-
duce similar results,

Method #1: MAP = 
DP + (1/3 (SP - DP))

Method #2: MAP = 
(2/3 DP) + (1/3 SP)

Method #3: MAP = 
((2*DP) + SP)/3

Important

The PASCO Blood 
Pressure Sensor is 
designed for educa-
tional use 

only

. It 

cannot be used for 
diagnosis.

Содержание PS-2207

Страница 1: ... the small cuff and PS 2209 contains the large cuff See the PASCO catalog or web site Included Equipment Part Number Blood Pressure Sensor with 3 available cuff sizes PS 2207 PS 2208 PS 2209 Blood Pressure Cuff with Bulb Valve 3 sizes available PS 2531 PS 2532 PS 2533 Recommended Items PASPORT Extension Cable PS 2500 PASCO Interface Catalog or web site Data Acquisition Software Catalog or web site...

Страница 2: ...blood on the artery walls when the ventricles relax and the heart s chambers fill with blood It is the bottom number of the blood pressure ratio Normal diastolic pressure for a male is approximately 80 mmHg and for females is approximately 70 mmHg Systolic and diastolic pressures are affected by various biological and environmental factors For example the salt in a person s diet can cause the kidn...

Страница 3: ... quick connector port on the sensor Push the quick connector onto the port and turn the connector clock wise until the connector clicks into place on the port 3 Connect the Blood Pressure Sensor to a PASPORT interface 4 If you will be using a computer connect the PASPORT interface to the com puter s USB port 5 Create Digits displays of pressure diastolic pressure and systolic pressure Your may als...

Страница 4: ...in data collection and then he or she should use the bulb to pump air into the cuff 3 Monitor the pressure in the Digits display 4 When the pressure reaches approximately 170 mmHg partner 2 should stop pumping and let go of the bulb 5 When the cuff is deflated press Stop to end data collection After the measure ment is finished release the remaining air in the cuff by pressing the push button rele...

Страница 5: ...blips small bumps This is your sys tolic pressure 3 Next identify when your data stops forming blips small bumps This is your diastolic pressure 4 You may need to change your sample rate in order to see the blips Identify when your data begins to have blips small bumps This is your systolic pressure Identify when your data stops forming blips small bumps This is your diastolic pressure You may nee...

Страница 6: ...elease rate is normally slower if the person s arm is larger Storage Disconnect the blood pressure cuff from the sensor for storage Average Blood Pressure Values Specifications Measurement Range Normal 120 80 or less Pre hypertension 120 80 to 140 90 High greater than 140 90 Very high greater than 180 110 Heart Rate Blood Pressure Gage Pressure Units beats per minute bpm millimeters of mercury mmH...

Страница 7: ... in the United States and or in other countries For more information visit www pasco com legal Windows is a registered trademark of Microsoft Corporation in the United States and or other countries Mac is trademark of Apple Com puter Inc registered in the U S and other countries Product End of Life Disposal Instructions This electronic product is subject to disposal and recycling regulations that ...

Отзывы: