Pasport PS-2121 Скачать руководство пользователя страница 4

Using the Colorimeter with DataStudio

By default, when you plug a Colorimeter into a PASPORT interface and click 
the 

Start 

button, DataStudio automatically displays the percent 

transmittance readings for each of the four colors (red, green, blue, orange) 
in a Graph display.  Any time you need to change the Colorimeter’s settings 
to read a different measurement, you must use the PASPORT Setup 
window in DataStudio.  To open the PASPORT Setup window, click the 

 button on the main toolbar.  Then click to place an “X” next to the 

desired measurement.  To  remove the setting, click on the “X” again.

Example

:  You want the Colorimeter to read absorbance instead of % 

transmittance for all four colors.  In the boxes next to the colors with the 
“transmittance” unit, click to remove the “X.”  Then click to place an “X” in 
each of the boxes next to the colors with the “Absorbance” unit.

1. Plug the Colorimeter into the PASPORT interface.  When the 

PASPORTAL window opens, select “Launch EZ-SCREEN.”

2. To begin collecting data, click the 

Start/Stop 

 button.

3. To end data collection, click the 

Start/Stop

 button.

4. To cycle through the different colors and measurement types, click 

the 

Colorimeter icon 

 in the bottom right corner of the screen.

5. To toggle between separate runs of data, click the 

Toggle

 

 

button.

Note:

  To simultaneously view all colors in one graph, you must use either 

DataStudio or DataStudio Lite.

Note: 

 Colorimeter options appear in the Setup window when you connect 

a Colorimeter to a USB link.  To view all Colorimeter measurement options, 
use the 

Maximize

 button in the Setup window. 

Using the Colorimeter with Xplorer

1. Turn on the Xplorer.
2. Plug the Colorimeter into the Xplorer. (Note: If you wish to perform 

calibration, you must have Xplorer connected to your computer. Use 
the procedure described on the reverse of this card.) 

3. To collect data, click the 

Start

 

 button.  The LED light flashes 

when Xplorer is storing data. (To stop, click the Start button again.)

4. To cycle through the measurements (transmittance or absorbance) for 

each color, click the 

Plus

 button (to more forward) or 

Minus

 button (to 

move backward.)

5. To view your data in DataStudio, connect the Xplorer cable to the USB 

port on your computer and to the USB port on the side of Xplorer.

6. At the DataStudio prompt, click 

Retrieve Now

.  DataStudio 

automatically displays the % transmittance readings for the four 
colors.  

Note: 

To view absorbance readings in DataStudio, you must open the Setup 

window and change the settings. See “Using the Colorimeter with DataStudio” 
on this card.   

Using the Colorimeter with the 

PASPORT EZscreen

Use the down 
arrow and the 

Plus

 

and 

Minus

 buttons 

to change the 
sample rate. 

012-08158C

CARD 4A

CARD 4B

Содержание PS-2121

Страница 1: ...smittance Resolution 0 1 transmittance Default sample rate 1 sample sec Maximum sample rate 5 samples sec Temp Range 5 40 C recommended 012 08158C 1 2 3 Colorimeter The recommended sample rate is 1 Hz for color and Beer s law studies If you plan to do reaction order studies use a faster sample rate The temperature range applies to both the temperature of the solution in the cuvette and the ambient...

Страница 2: ...nd forth to evenly mix the solution without creating bubbles 3 Insert the cuvette containing distilled water into the cuvette holder and tightly close the lid 4 Plug the Colorimeter into the PASPORT interface launch DataStudio and click the Start button to begin collecting data The default graph displays the transmittance readings for each of the four possible wavelength settings Click the Stop bu...

Страница 3: ...he sample 4 To check your calibration reinsert the cuvette close the lid and click the Start button in DataStudio All color readings should approximate 100 transmittance 1 2 3 Experiment Suggestions Performing reaction rate enzymatic and bacterial growth studies Monitoring the concentration of substances in water or other solutions 012 08158C Orienting the Cuvette Optional To minimize reading vari...

Страница 4: ...e runs of data click the Toggle button Note To simultaneously view all colors in one graph you must use either DataStudio or DataStudio Lite Note Colorimeter options appear in the Setup window when you connect a Colorimeter to a USB link To view all Colorimeter measurement options use the Maximize button in the Setup window Using the Colorimeter with Xplorer 1 Turn on the Xplorer 2 Plug the Colori...

Отзывы: