Pasport PS-2121 Скачать руководство пользователя страница 3

Calibration Procedure 

Sample Preparation

Calibration of the Colorimeter is not always required, but is recommended for 
better accuracy.  You can calibrate the colorimeter when it is connected to either 
an Xplorer or a USB link.  All calibration readings are stored in flash memory 
inside the PS-2121.  When you unplug the sensor and reconnect it, the sensor 
retains the last calibration reading.

1. Fill the cuvette with at least 6 ml of sample. Screw the cap on securely.
2. Gently rock the cuvette back and forth to distribute the solute molecules 

equally throughout the solution.  

Do not shake the cuvette or allow air 

bubbles to enter.

3. Handle the cuvette by the cap and wipe the outside glass lens clean with 

a non-abrasive cleaning tissue.  

Avoid touching the glass with your 

fingers!

4.  (Optional): Place one of the provided labels on the cap of the cuvette.  (See 

“Orienting the Cuvette” below.)  

Before taking measurements, always do the following:

  

CAUTION:

  To avoid errors when calibrating, do not use the Colorimeter 

around bright lights or with turbid solutions.  Always close the lid tightly before 
taking measurements. For accurate readings, fill the cuvette to the lid with 
sample.

1. To calibrate for 100% transmittance, insert a cuvette filled with distilled 

water into the cuvette holder and tightly close the lid.

2. Press the green 

Calibration

 button on the sensor.  The light 

illuminates to indicate calibration is in progress.

3. Wait for the light to turn off and then remove the sample.
4. To check your calibration, reinsert the cuvette, close the lid, and click 

the 

Start 

button in DataStudio.  All color readings should approximate 

100% transmittance.

1

2

3

Experiment Suggestions

• Performing reaction rate, enzymatic, and bacterial growth studies
• Monitoring the concentration of substances in water or other solutions

012-08158C

Orienting the Cuvette (Optional)

To minimize reading variance due to natural differences or imperfections in the 
glass of the cuvette, do the following for each cuvette you plan to use:
1. Insert the cuvette containing distilled water or other solution into the 

cuvette holder.

2. Put a piece of black cloth over your hand and the cuvette (so that no 

light seeps through). Click the 

Start 

button in DataStudio. 

3. With your covered hand, rotate the cuvette while observing the reading 

in DataStudio. At the highest transmittance reading, click the 

Stop

 

button. Do not rotate the cuvette any further.

4. With the cuvette still in the holder, place one of the provided arrow labels 

on the cap, with the arrow pointing toward the screw on the holder. (For 
subsequent measurements, always align the arrow with the screw on 
the holder. Do not switch caps between cuvettes!)  

Note:

  While in calibration mode, please do not start collecting data. The 

PS-2121 Colorimeter cannot communicate during this time and an error 
will occur. 

A blinking red light on the Calibration button means either (a): The dark 
count is too high; stray light is entering the cuvette holder or (b): The 
sensor measurement is out of range. The light turns off when the reading 
is within the normal range. 

800-772-8700  •  916-786-3800  •  [email protected]  •  www.pasco.com

®

CARD 3A

CARD 3B

Содержание PS-2121

Страница 1: ...smittance Resolution 0 1 transmittance Default sample rate 1 sample sec Maximum sample rate 5 samples sec Temp Range 5 40 C recommended 012 08158C 1 2 3 Colorimeter The recommended sample rate is 1 Hz for color and Beer s law studies If you plan to do reaction order studies use a faster sample rate The temperature range applies to both the temperature of the solution in the cuvette and the ambient...

Страница 2: ...nd forth to evenly mix the solution without creating bubbles 3 Insert the cuvette containing distilled water into the cuvette holder and tightly close the lid 4 Plug the Colorimeter into the PASPORT interface launch DataStudio and click the Start button to begin collecting data The default graph displays the transmittance readings for each of the four possible wavelength settings Click the Stop bu...

Страница 3: ...he sample 4 To check your calibration reinsert the cuvette close the lid and click the Start button in DataStudio All color readings should approximate 100 transmittance 1 2 3 Experiment Suggestions Performing reaction rate enzymatic and bacterial growth studies Monitoring the concentration of substances in water or other solutions 012 08158C Orienting the Cuvette Optional To minimize reading vari...

Страница 4: ...e runs of data click the Toggle button Note To simultaneously view all colors in one graph you must use either DataStudio or DataStudio Lite Note Colorimeter options appear in the Setup window when you connect a Colorimeter to a USB link To view all Colorimeter measurement options use the Maximize button in the Setup window Using the Colorimeter with Xplorer 1 Turn on the Xplorer 2 Plug the Colori...

Отзывы: