background image

®

Wireless Load Cell ±50N

PS

-

3216

6

013-15734A

clamp the fixture at the edge of a table and hang a mass 

from the sensor as shown. 

Note that the hanging mass applies tension to the LOAD 

side of the Wireless Load Cell; therefore the known force 

that you enter into the software should be a negative 

value. For example, if the mass is 0.5 kg, the applied 

force is -4.9 N.

Direction of Acceleration Sensitivity

The acceleration sensing units inside the sensor are 

oriented so that the lines of greatest sensitivity follow the 

three X-Y-Z arrows indicating the direction of 

acceleration. Note that the software calculates the mag

-

nitude of the resultant acceleration based on the X-Y-Z 

values. 

X-Y-Z Orientation

In this orientation of the sensor, the X-axis direction is 

‘horizontal’, the Y-axis direction is ‘vertical’ and, and the 

Z-axis direction is ‘perpendicular’ to the label on the front 

of the sensor.

Note that a compression force (LOAD side pushed 

toward FIXED side) is considered positive and a tension 

force (LOAD side is pulled away from FIXED side) is 

considered negative in direction.

Test the Sensor’s Output Using Earth’s Gravity

Start the PASCO Data Collection Software. If the Wire

-

less Load Cell is arranged so that the X-axis on the top 

of the sensor is vertical and the LOAD side of the Wire

-

less Load Cell is at the top, the Acceleration X-axis 

reading is 9.8 m/s². If the sensor is turned so the X-axis 

is vertical but the LOAD side is at the bottom, the 

Acceleration X-axis reading is -9.8 m/s².

2

2

2

R

X

Y

Z

Arrows indicate direction of acceleration

Z

X

Y

9.8 m/s²

-9.8 m/s²

Содержание PS-3216

Страница 1: ...The PASCO Data Collection Software can show the compression or tension forces the accelera tion in X Y and Z and the resultant acceleration The sensor uses a micro electromechanical system MEMS device...

Страница 2: ...he sensor off press and hold the ON OFF button for a moment until the Battery Status LED shines red The Wireless Load Cell puts itself to sleep after several minutes of inactivity if not connected and...

Страница 3: ...tton All of the Force and Acceleration measurements will be set to zero To Zero Force or Acceleration Another option is to set either the Force measurement or the Acceleration measurements to zero Sta...

Страница 4: ...option is to use the Common Rate button Click the Common Rate button in the Control palette below the workbook to open the Common Rate menu Select Wireless Force Sensor from the menu and then click th...

Страница 5: ...th the PASCO Structures Systems such as the ME 3581 Building Better Bridges Kit The sensor can measure compression and tension forces in any member of the PASCO Structures Systems Use thumbscrews to m...

Страница 6: ...Y Z Orientation In this orientation of the sensor the X axis direction is horizontal the Y axis direction is vertical and and the Z axis direction is perpendicular to the label on the front of the sen...

Страница 7: ...rtant to making the sensor simple and always ready to use so all of the PASCO wireless products are designed for long battery life For example the sensor turns itself off after a few minutes of inacti...

Страница 8: ...ence and 2 this device must accept any interference received including interference that may cause undesired operation CE Statement This device has been tested and found to comply with the essential r...

Страница 9: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Страница 10: ...this example click 2 point 6 Click Next The Calibrate Sensor Enter Values screen opens 7 Hang the 1 kilogram mass on the calibration fixture 8 The gravitational force on the mass is pulling in the neg...

Страница 11: ...linear equation that transforms the raw input data into the calibrated output data The linear function is of the form Raw Input Slope x Calibrated Output Offset Or Calibrated Output Raw Input Offset...

Страница 12: ...e shifts so that it intersects the new point but its slope does not change Offset calibration is usually used to make one sensor agree with another sensor Due to normal variation among probes a second...

Отзывы: