manualshive.com logo in svg
background image

16. Click "OK" to close the Calibrate Sensor window, then

click anywhere in the Graph display to close the "Configure
sensors and measurements" menu.

About the Calibration

Calibration information is stored in the Wireless Drop Counter
and can be used by both PASCO Capstone and SPARKvue. The
information can be changed as needed. Note that if you use a
different liquid as the titrant, or if you change the drop rate or
drop size by adjusting the stopcock valves on the Drop
Dispenser, you will need to recalibrate.

Titration Setup

Figure 2. The titration setup, with the Wireless Drop Counter and syringe
holder mounted to a support rod.

 Syringe with 50 mL of 0.1 M NaOH

 Syringe Holder

 Wireless pH Sensor

 Stopcocks

 Direct Connect pH Probe

 Drop Tip

 Wireless Drop Counter

 150 mL Beaker with 100 mL of 0.005 M HCl

 Micro Stir Bar

 Magnetic Stirrer

Assemble the Titration Setup

1. Mount the Wireless Drop Counter on a support rod.

2. Attach a pH Probe to a Wireless pH Sensor (PS-3204).

Place the pH probe into the larger of the two sensor ports
on the Wireless Drop Counter.

3. Mount the Micro Stir Bar to the end of the pH Probe.

4. Arrange the magnetic stirrer and a 150 mL beaker with 100

mL of 0.005 M HCl, such that the end of the pH Probe and
the Micro Stir Bar are submerged in the liquid in the beaker.

5. Mount the Drop Dispenser above the Wireless Drop

Counter and position it so that the drop tip is placed over
the rectangular window of the Drop Counter. Fill the syringe
with 50 mL of 0.1 M NaOH.

Acid-Base Titration

1. Set up the equipment as shown in Figure 2, including

consumables.

2. Start PASCO Capstone or SPARKvue. Connect the

Wireless Drop Counter and a Wireless pH Sensor to the
software.

3. Use the software to create a graph display of pH (vertical

axis) vs. Fluid Volume (horizontal axis).

4. Start the magnetic stirrer. Check that the micro stir bar is

able to rotate.

5. In the software, click "Start" (SPARKvue) or "Record"

(Capstone) to begin recording data.

6. Adjust the stopcock valves on the Drop Dispenser so that

the titrant drops fall at about 1 drop per second.

7. Observe the data on the pH versus Fluid Volume graph.

After the equivalence point is reached, continue collecting
data until the pH curve flattens.

8. Close the stopcock valve and stop data recording.

Other Titrations

Titrations with different solutions can be performed using the
same equipment and procedure. Other probes, such as
conductivity or ORP, can be used in place of the pH probe.

Product Guide | 012-16026B

5

Содержание PS-3214

Страница 1: ...rt It includes a rechargeable battery and a Micro USB Cable It measures the number of drops of fluid that fall through the rectangular opening drop window of the sensor and converts the drops into a volume It is typically used in conjunction with a pH sensor and other equipment to perform a titration Data from the Wireless Drop Counter and other sensors are recorded and displayed using software su...

Страница 2: ...utton until the lights turn on 2 Open PASCO Capstone 3 Click Hardware Setup 4 Select the sensor that matches its device ID A notification will appear if a firmware update is available Click Yes to update the firmware 5 Close Capstone ON OFF and Sleep Information Before turning the sensor on make sure that nothing is blocking the drop window To turn the sensor on press and hold the ON button When t...

Страница 3: ... adjust the valve on the bottom stopcock so that the drops fall at a rate of about 1 drop per second Close the top valve after the rate is achieved Discard the titrant in the beaker as directed by the instructor 4 Mount the Wireless Drop Counter to the same support rod as the Drop Dispenser between the Dispenser and the base 5 Align the drop tip with the rectangular window in the Wireless Drop Cou...

Страница 4: ... Sensor Data b Turn on the Wireless Drop Counter c From the list of available Bluetooth devices select the device matching the Device ID number XXX XXX printed on your sensor 8 Select a Template from the options on the right to enter the Experiment window 9 Select the Hardware Setup icon in the bottom right corner This window shows the name of all connected sensors and selected measurement choices...

Страница 5: ...ic stirrer and a 150 mL beaker with 100 mL of 0 005 M HCl such that the end of the pH Probe and the Micro Stir Bar are submerged in the liquid in the beaker 5 Mount the Drop Dispenser above the Wireless Drop Counter and position it so that the drop tip is placed over the rectangular window of the Drop Counter Fill the syringe with 50 mL of 0 1 M NaOH Acid Base Titration 1 Set up the equipment as s...

Страница 6: ...terproof Do not immerse the sensor in liquid If the inside surfaces of the drop window need to be cleaned use a piece of lens paper or a lint free cloth and water or a glass cleaner Tips Except for opening and closing the stopcocks do not touch the syringe of the Drop Dispenser during the titration otherwise the drop size may change significantly The drop rate must remain approximately constant fo...

Страница 7: ...al requirements and other relevant provisions of the applicable EU Directives FCC statement This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Battery disposal Batteries contain chemic...

Отзывы: