manualshive.com logo in svg
background image

10. In Capstone, highlight the number in the "Standard Value"

text box in Step 4. Enter the exact volume of titrant that you
collected in the graduated cylinder (5 mL in this example),
then click the button "Set Current Value to Standard Value".

11. Step 5 automatically opens, showing "New Calibration" in

the lower half. In this example, Capstone counted 10 drops
and the volume was 5.0 mL, so the "New Calibration"
shows "Slope" equal to "0.500 mL/drops".

12. Click "Finish" to return to Step 1, then click the "Calibration"

icon in the Tools panel again to close the Calibration
window.

Calibration using SPARKvue

1. Attach the Drop Dispenser to a support rod if you have not

already done so.

2. Fill the syringe with titrant. Put a beaker under the Drop

Dispenser to catch the titrant.

3. Open both stopcock valves and adjust the valve on the

bottom stopcock so that the drops fall at a rate of about 1
drop per second. Close the top valve after the rate is
achieved. Discard the titrant in the beaker as directed by
your instructor.

4. Mount the Wireless Drop Counter to the same support rod

as the Drop Dispenser, between the Dispenser and the
base.

5. Align the drop tip with the rectangular window of the

Wireless Drop Counter. Place a graduated 10 mL cylinder
under the Drop Counter window so it can catch drops from
the drop tip and give you a measure of the volume.

6. Start SPARKvue. The "choose a path" window opens.

7. Connect the Wireless Drop Counter to SPARKvue using

the following steps:

a. In SPARKvue, click or touch "Sensor Data".

b. Turn on the Wireless Drop Counter.

c. From the list of available Bluetooth devices, select the

device matching the Device ID number (XXX-XXX)
printed on your sensor.

8. Select a Template from the options on the right to enter the

Experiment window.

9. Select the Hardware Setup 

 icon in the bottom right

corner. This window shows the name of all connected
sensors and selected measurement choices. Note that
each measurement has a gear-shaped "Edit Data

Properties" icon; in this case, the Fluid Volume
measurement also has a "Calibrate measurement" icon.

10. Select the "Calibrate Measurement" icon next to "Fluid

Volume" to open the Calibrate Sensor window.

11. In the Calibrate Sensor window, click "Continue" once you

have ensured the Drop Dispenser and Wireless Drop
Counter are properly in position. A screen containing a text
area for "Standard Value" and a display area for "Sensor
Value" will open.

12. Open the top stopcock valve on the Drop Dispenser to

allow titrant to begin dripping into the cylinder. Note that
SPARKvue will automatically begin recording the number of
drops (you do not need to click "Record").

13. After about 5 mL of titrant has been collected in the

graduated cylinder, close the bottom stopcock. Carefully
measure the exact volume of titrant in the graduated
cylinder.

14. Highlight the number in the "Standard Value" text box and

enter the exact volume of titrant that you collected in the
graduated cylinder (5.0 mL in this example), then click "Set
Calibration".

15. Optional: Click "New Calibration" to see the Slope and

Offset values. In this example, the Slope is "2.0000
drops/mL" because 10 drops were counted and the volume
of the titrant entered is 5 mL.

Wireless Drop Counter | PS-3214

4

Содержание PS-3214

Страница 1: ...rt It includes a rechargeable battery and a Micro USB Cable It measures the number of drops of fluid that fall through the rectangular opening drop window of the sensor and converts the drops into a volume It is typically used in conjunction with a pH sensor and other equipment to perform a titration Data from the Wireless Drop Counter and other sensors are recorded and displayed using software su...

Страница 2: ...utton until the lights turn on 2 Open PASCO Capstone 3 Click Hardware Setup 4 Select the sensor that matches its device ID A notification will appear if a firmware update is available Click Yes to update the firmware 5 Close Capstone ON OFF and Sleep Information Before turning the sensor on make sure that nothing is blocking the drop window To turn the sensor on press and hold the ON button When t...

Страница 3: ... adjust the valve on the bottom stopcock so that the drops fall at a rate of about 1 drop per second Close the top valve after the rate is achieved Discard the titrant in the beaker as directed by the instructor 4 Mount the Wireless Drop Counter to the same support rod as the Drop Dispenser between the Dispenser and the base 5 Align the drop tip with the rectangular window in the Wireless Drop Cou...

Страница 4: ... Sensor Data b Turn on the Wireless Drop Counter c From the list of available Bluetooth devices select the device matching the Device ID number XXX XXX printed on your sensor 8 Select a Template from the options on the right to enter the Experiment window 9 Select the Hardware Setup icon in the bottom right corner This window shows the name of all connected sensors and selected measurement choices...

Страница 5: ...ic stirrer and a 150 mL beaker with 100 mL of 0 005 M HCl such that the end of the pH Probe and the Micro Stir Bar are submerged in the liquid in the beaker 5 Mount the Drop Dispenser above the Wireless Drop Counter and position it so that the drop tip is placed over the rectangular window of the Drop Counter Fill the syringe with 50 mL of 0 1 M NaOH Acid Base Titration 1 Set up the equipment as s...

Страница 6: ...terproof Do not immerse the sensor in liquid If the inside surfaces of the drop window need to be cleaned use a piece of lens paper or a lint free cloth and water or a glass cleaner Tips Except for opening and closing the stopcocks do not touch the syringe of the Drop Dispenser during the titration otherwise the drop size may change significantly The drop rate must remain approximately constant fo...

Страница 7: ...al requirements and other relevant provisions of the applicable EU Directives FCC statement This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Battery disposal Batteries contain chemic...

Отзывы: