Model  No.PS-2195

 Notes on the Experiment

9

Notes on the Experiment

If bubbles form inside the probe, the conductivity reading will be reduced because the 
bubbles will form an insulating later on one or both of the electrodes. One way to 
eliminate the bubbles is to tap the probe. Another way is to increase the speed of the 
magnetic stirrer to allow more solution to flow through the probe.

If time is limited, prepare the solutions before the period begins.

Data Table

Data Analysis

The table lists typical experimental results. In general, ionic salts at low to moderate 
concentrations have a temperature dependence of 2% per degree at 25 °C. Acids, 
bases, and concentrated salt solutions have somewhat lower values, typically 1.5% 
per °C. In contrast, ultra pure water has a much larger value; 5.2% per °C.

Questions

1.

The conductivity increases linearly with temperature over the observed tempera-
ture range.

2.

The slopes are approximately equal for all the solutions.

3.

Temperature, concentration, and solubility will affect the conductivity of a solu-
tion.

Solution

percent change/ °C at 25 °C

0.1% NaCl (1000 ppm)

2.1

0.4% NaCl (4000 ppm)

2.0

0.005 M NaOH (1000 ppm)

1.9

Содержание PasPort PS-2195

Страница 1: ...linity Sensor PS 2195 Salinity Sensor Probe 10 X Conductivity Temperature 699 11064 Recommended Items PASPORT Extension Cable PS 2500 PASCO Interface Catalog or web site Data Acquisition Software Cata...

Страница 2: ...ride In the 1970 s salinity was redefined as the conductivity ratio of a water sample to a standard potassium chloride KCl solution Usage One use for the sensor is to explore the salinity of local wat...

Страница 3: ...b site at www pasco com for information 1 When you connect the Salin ity Sensor to the computer through a PASPORT inter face the PASPortal window will launch automatically if DataStudio is not already...

Страница 4: ...ou will need reagent grade sodium chloride salt a liter of distilled or deionized water a stir rod and a container with accurate volume markings Pour 500 milliliters mL of distilled water into the con...

Страница 5: ...on the specific dissolved solids and can vary between 0 40 and 0 96 depending on the dissolved solids A value of 0 65 is used as an approximation if the dissolved solids are not known As an example TD...

Страница 6: ...f PASCO scientific in the United States and or in other countries For more information visit www pasco com legal Windows is a registered trademark of Microsoft Corporation in the United States and or...

Страница 7: ...g more distilled or deionized water until the volume is 200 ml Prepare a 0 005 M sodium hydroxide NaOH solution by dissolving 200 mg of NaOH in 500 ml of distilled or deionized water and then adding d...

Страница 8: ...Remove the Salinity Sensor Probe from the first sample Rinse the end of the probe with distilled water 9 Repeat the data collection process with the other two samples Data Analysis 1 Autoscale the gra...

Страница 9: ...Analysis The table lists typical experimental results In general ionic salts at low to moderate concentrations have a temperature dependence of 2 per degree at 25 C Acids bases and concentrated salt s...

Страница 10: ...Salinity Sensor Notes on the Experiment 10...

Отзывы: