background image

M o d e l   N o .   P S - 2 1 7 0

  S e n s o r ,   I n t e r f a c e ,   a n d   S o f t w a r e   S e t u p

3

012-09523B

Computer Setup

Using PASCO Capstone

Connect a PASPORT-compatible interface to the computer and start the PASCO Capstone software. 
(Refer to the documentation that came with your PASPORT interface for instructions on connecting 
it to your computer.) In Capstone, click the Record button to begin data collection; click the Stop 
button to stop data collection.

Note that the sensor may collect data for probes that are not 
connected. To hide measurements that you do not need, 
click the Data Summary icon (

) in the Tools palette to 

open the Sensor Data Summary panel. Click the Visibility 
icon (

) to view the menu of parameters. Check or 

un-check the measurements as needed. Click the Data Sum-
mary icon again to close the panel.

In the Workbook page, click one of the configured display 
combinations, or drag icons from the Displays palette into the Workbook page. In a display, 
click the “<Select Measurement>” menu to view the sensor parameter choices. Click a param-
eter in the menu to set the measurement for that display. 

Using SPARKvue HD

Connect a PASPORT-compatible interface to the computer and start the 
SPARKvue HD software. (Refer to the documentation that came with your 
PASPORT interface for instructions on connecting it to your computer.) In 
SPARKvue HD, the opening screen shows the list of measurements for the sen-
sor. For example, click “Temperature” and then click “Show” to open a Graph 
display screen.

In the Graph display screen, click the green Record button (

) to begin record-

ing data. Click the red Stop button (

) to stop recording data.

Wireless Setup

Using SPARKvue HD

Use the SPARKvue HD software to pair a mobile device such as a tablet or 
smart phone to a PASCO wireless interface such as the AirLink2 or SPARKlink 
Air. (Refer to the documentation that came with your PASCO wireless interface 
for instructions on connecting it to your mobile device.) 

Connect the sensor to the wireless interface. In SPARKvue HD, the opening 
screen shows the list of measurements for the sensor. For example, touch 
“Temperature” and then touch “Show” to open a Graph display screen.

In the Graph display screen, touch the green Record button (

) to begin 

recording data. Touch the red Stop button (

) to stop recording data.

Record

Stop

Touch 

here.

Then touch 

here.

Содержание PASPort PS-2170

Страница 1: ...yurethane Tubing 60 cm 0 125 inch ID 6 Quick release Connectors qty 4 Hose and Coupling Kit 7 Tubing Connectors qty 4 8 Syringe 60 cc 9 Alligator Clip with Insulator qty 2 each Polyurethane Tubing blu...

Страница 2: ...RP port BNC port Align the grooves of the BNC connector on the probe with the pins on the BNC port Push the connector onto the port and turn the connector clockwise until it locks in place If you plan...

Страница 3: ...easurement for that display Using SPARKvue HD Connect a PASPORT compatible interface to the computer and start the SPARKvue HD software Refer to the documentation that came with your PASPORT interface...

Страница 4: ...measurement can be displayed in units of C F or K You can use the included stainless steel probe or a different probe such as the Fast response Temperature Probe PS 2135 or Skin Surface Temperature Pr...

Страница 5: ...Electrode Maintenance and Storage Cleaning If the pH electrode becomes contaminated use one of these methods to clean and restore it After any of these procedures rinse the electrode with deionized w...

Страница 6: ...tors to connect the tubing to larger diameter tubing or to a rubber stopper with a hole Note Do not allow liquid to enter the pressure port Voltage Use the voltage component of the multi sensor to mea...

Страница 7: ...Model No PS 2170 Specifications 7 012 09523B Voltage Range 10 V to 10 V Accuracy 0 1 V Resolution 0 04 mV Voltage Protection up to 240 V Input Resistance 2 M...

Страница 8: ...ept Move to the third step 3 The third step in the panel is to choose the type of calibration Two Standards 2 point is the default Click Next 4 The fourth step in the panel is to calibrate the first p...

Страница 9: ...he Graph display screen to open the Experiment Tools menu In the Experiment Tools menu touch Calibrate Sensor The screen will show Calibrate Sensor Select Measure ment The Sensor Chemistry Measurement...

Страница 10: ...be in distilled water 5 For Calibration Point 2 touch the down arrow to scroll down to the bottom of the screen Place the probe s sensor into the second known standard such as the pH 7 buffer solution...

Страница 11: ...s pH Probe into the first known standard pH 4 buf fer solution Wait a few moments and then press F3 H to Read Pt 1 5 Use the down arrow to scroll down to the second calibration point Highlight Pt 2 mV...

Страница 12: ...products or service names are or may be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Teflon i...

Отзывы: